Product Name: Total Nitrogen Water Quality Online Automatic Monitor Product Model: TY-TN Product Overview The total nitrogen water quality online automatic monitor is developed by Tianjin Tongyang Technology Development Co., Ltd. for many years ..

Product Details


The total nitrogen water quality online automatic monitoring instrument is a fully automatic online monitoring instrument launched by Tianjin Tongyang Technology Development Co., Ltd. based on years of research on water quality analysis products. After high-temperature digestion of the sample, all other valence state nitrogen in the sample is oxidized to nitrate nitrogen. Under acidic conditions, the tested sample reacts with the color developer to generate stable soluble compounds. Within the measurement range of the instrument, the degree of color change is proportional to the total nitrogen concentration in the sample. Therefore, by measuring the degree of color change using the photoelectric colorimetric method, the concentration of total nitrogen in the sample can be calculated


Technical Features


lThe dual beam photometer measurement eliminates the interference of external factors on the measurement results;

Optoelectronic precise quantification, achieving reagent miniaturization;

L has functions such as automatic calibration and automatic cleaning;

Accurate temperature control technology overcomes the influence of environmental temperature;

lIntelligent design for fault self diagnosis makes instrument management and maintenance simple and convenient.


Technical Specifications



project

technical indicators

Measuring principle

Potassium persulfate spectrophotometric method

measurement range

050)mg/LThe range can be customized according to customer requirements

repetitiveness

10%

zero drift

±5%

range drift

±10%F.S

Minimum measurement cycle

45min

voltage

AC220V±10%

insulation resistance

5MΩ

communication interface

RS485/RS232/4-20mA/1-5V

Installation method

cabinet-style


Scope of application


Applied to monitoring pollution source discharge outlets, municipal sewage, industrial sewage, industrial production water, etc.