Product Name: Benzene series online gas chromatograph Product Model: TY-VOC500 (FID) Product Overview A. Instrument Principle ..
A. Instrument principle
The instrument adopts a combination of dedicated chromatography columns, center cutting and blowback technology, and hydrogen flame ionization detector(FID)After the detection of benzene derivatives and the collection of samples in the quantitative loop, the injection valve is switched. Under the drive of the carrier gas, the samples enter the chromatographic column for separation. Among them, benzene, toluene, xylene, and ethylbenzene enter the analytical column for separation after passing through the pre columnFIDThe detector detected that after switching the valve position, the heavy hydrocarbon components were blown back and vented.
B. Instrument features
Using center cutting and back blowing technology to directly measure benzene, toluene, and xylene;
Using a dedicated chromatographic column combination, with no residue and high sensitivity;
The instrument has a power on self-test function, gas cut-off protection function, automatic restart function upon power failure, and alarm function to ensure system safety and stability;
The FID detector has automatic ignition function and wide range output, with a linear range 10-7;
Using automatic electronic flow control technology(EPC)Control carrier gas, air, and hydrogen with high precision(0.01psi)Good repeatability and reproducibility;
The core components are all imported brands, with high reliability and long service life;
monitoring parameters
Benzene, toluene, xylene, meta xylene, para xylene, and isopropylbenzene
Technical Specifications
| project | technical indicators |
| detection capability | Benzene, toluene, xylene, meta xylene, para xylene, isopropylbenzene |
| range | Benzene (0.05-10000) ppm; Toluene (0.05-10000) ppm; Xylene (0.1-10000) ppm |
| detector | Hydrogen flame ionization detector (FID) |
| detection limit | ≤ 0.03ppm (benzene); |
| repetitiveness | RSD ≤ 3% (benzene) |
| Analysis cycle | Single channel ≤ 15min, dual channel ≤ 8min |
| Power supply | < 400W,220V AC/50Hz |
| working environment | Temperature: (-10~50) ℃, Humidity: (10%~90%) RH |
| Air source requirements | Carrier gas: high-purity nitrogen (≥ 99.999%); Combustion gas: high-purity hydrogen gas (≥ 99.999%); Auxiliary gas: Zero grade air (hydrocarbons<20ppb) |
| output | 4-20mA、RS232/RS485、 Ethernet |
| size | 19 inch standard chassis, 5U |
Scope of application
Discharge outlets for petrochemical, spraying, printing, etc
Online monitoring at the factory boundary