Welcome Customer !

Membership

Help

Suzhou Yucheng Instrument Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products

Suzhou Yucheng Instrument Co., Ltd

  • E-mail

    2524300166@qq.com

  • Phone

    18015580277

  • Address

    No. 55 Dongfu Road, Suzhou Industrial Park

Contact Now

AWA6256B+Environmental Vibration Analyzer

NegotiableUpdate on 04/04
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The AWA6256B+environmental vibration analyzer consists of an environmental vibration accelerometer, a host, and environmental vibration measurement and analysis software, mainly used for environmental vibration measurement. Environmental vibration can comply with both ISO 8041:1990 and GB/T 23716-2009 (ISO 8041:2005) standards, meet the requirements for instruments in the current GB/T 10071-1988 standard, and also meet the requirements for environmental vibration measurement instruments in the revised standard.
Product Details

AWA6256B+Environmental Vibration AnalyzerComposed of an environmental vibration accelerometer, host, and environmental vibration measurement and analysis software, mainly used for environmental vibration measurement. Environmental vibration can comply with both ISO 8041:1990 and GB/T 23716-2009 (ISO 8041:2005) standards, meet the requirements for instruments in the current GB/T 10071-1988 standard, and also meet the requirements for environmental vibration measurement instruments in the revised standard.

AWA6256B+Environmental Vibration Analyzertechnical parameters

performance metrics Environmental vibration measurement Human body vibration measurement Low frequency 1/3 vibration measurement
execution standard ISO 8041:1990

ISO 8041:2005

GB/T 23716-2009

IEC 61260:1995
sensor AWA14400 environmental vibration accelerometer sensitivity: 40 mV/m · s-2Quality: 550g.
frequency range

1 Hz~63 Hz±1 dB

1 Hz~80 Hz±2 dB

Wb/Wc/Wd/We/Wj/Wk:

(0.63~63)Hz; ±1 dB

(0.50~125)Hz; ±2 dB

Wm:

(1.25~63)Hz; ±1 dB

(0.63~125)Hz; ±2 dB

0.63 Hz~250 Hz; ±1 dB

0.315 Hz~250 Hz; ±2 dB

measurement range

48 dB to 158 dB (with 10 dB)-6m/s2For reference)

Frequency weighting Parallel W.B.z (whole body vertical, abbreviated as z weighting) and W.B.x-y (whole body horizontal, abbreviated as x weighting) Parallel (simultaneous) Wb, Wc, Wd, We, Wj, Wk, Wm AP, Wk, Wz, Wu (user-defined)
Linear range Greater than 90 dB
sampling frequency 750 Hz
A/D bits 24 digits
data store 64 kB of FLASH can store 128 sets of measurement results
output interface RS232, Can be connected to a micro printer to print measurement results, and can also import data into a computer
working power supply

6 LR6 alkaline or rechargeable batteries, can be used continuously for more than 16 hours, and can also be powered by a 5V external power source

working environment

Working temperature: -10 ℃~50 ℃

Storage temperature: -20 ℃~70 ℃