The HT-LA475 wall mounted oxygen analyzer is widely used for online analysis of oxygen content in industries such as protective atmosphere furnaces, high-temperature sintering furnaces such as magnetic materials and brazing furnaces, protective gas furnaces for electronic semiconductors, and glass process requirements.

Product Details

HT-LA475Wall mounted oxygen analyzer

Technical Specifications:

  • Measuring principlefuel cell
  • display mode320 × 240 dot matrix LCD
  • measurementscope0~10/100/1000PPm O2(Trace)

010%/25%/100%O2(Constant)

  • measurement accuracy0~1000PPm/10%/25%/100%±1%FS

0~100PPm±2% FS

0~10PPm±5%FS

  • smallresolution0.1PPm
  • repetitiveness±1%FS
  • Response time:T9060S(Trace)

T9040S(Constant)

  • outputinterface4-20mA.DC (non isolated output, load resistance)500Ω

05V. DC (non isolated output, load resistance>10K)Ω

1Programmable dry contact type passive alarm outputLarge contact capacityAC220V/2A

  • contact methodRS232 (default) or RS485
  • workpower supplyAC220V±10%50/60Hzpower consumption40VA
  • environmenttemperature-10~+50
  • storeambient humidity:90%RH, non condensing
  • workambient humidity:100%RH, non condensing
  • Sample gas temperature: 0~40
  • Sample gas flow rate: 400~600mL/min
  • Sampling method: suction orThrough type(Choose one of the two)
  • sample gasStressMicro positive pressure, micro negative pressure, or atmospheric pressureexhaust type

0.05Mpaentrancegauge pressure0.35MpaStable pressure airAtmosphereThrough type

  • Exhaust pressureFree emptying (under safe conditions)
  • Background gasN2inertiaAcidity, reducibility, etcgas
  • Specifications and Dimensions400mm×300mm×200mmH×W×D
  • sensorlifespan24amoon(Under normal usage conditions)
  • air circuit interfaceNPT 1/8insidethread
  • instrument weightNet weight17.5Kg
  • Installation methodwall-mounted

Instrument features:

  • The function of switching between Chinese and English menus is intuitive and convenient to operate;
  • Adopting imported fuel cell sensors, it has the characteristics of fast response speed, high measurement accuracy, long calibration cycle, and strong resistance;
  • Adopting a protective air path of * effectively protects the sensor and avoids the sensor's lifespan being affected by prolonged exposure to air;
  • Built in sensor protection device and temperature compensation sensor ensure the service life of the sensor while reducing the impact of changes in sample gas temperature and environment on measurement accuracy;
  • Calibration can be done at any point within the entire range, making calibration simple and convenient;
  • canright5 of them通道Sampling of gasMeasurement;
  • temperatureAnd pressurecompensationEliminate temperature and pressure effectsThe impact of measurement accuracy;
  • 40 thousandarticlehistoryAutomatic data storage function, and can be used at any time本地Refer to historical data;
  • Built in * sampling pump, long lifespan, reliable operation(Optional)
  • wall-mountedInstallation is simple and easy to maintain
  • According to user needs, RS232 (default) or RS485 data communication interfaces can be optionally configuredCan be used with computersOr other digital communication devices directlyPerform one-way or two-way communication.

Application scenarios:

Widely used for online analysis of oxygen content in industries such as protective gas furnaces for high-temperature sintering furnaces such as magnetic materials and brazing furnaces, protective gas furnaces for electronic semiconductors, and glass process requirements.

Ordering Instructions(Please specify when placing an order)

lmeasurement range

lTested gas pressure

lBackground gas composition

HT-LA475Wall mounted oxygen analyzer