Walk in salt spray chamber/constant temperature and humidity chamber

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The salt spray corrosion environment simulation laboratory can simulate typical environmental conditions in nature at any time without being limited by time and season. Putting the product into an environmental simulation laboratory for environmental simulation testing, replacing the usage environment with a simulated environment, can save a lot of manpower and material resources, and save development costs. At the same time, environmental simulation testing is based on the product's usage environment conditions, controlling the product's usage conditions. The testing has an "acceleration" property, which can shorten the product's testing cycle.

Product Details

name

Main technical parameters

salt spray chamber

Salt spray box (quantity: 1 unit)

1. Main technical parameters

1.1Studio volume: 2.5 cubic meters;

1.2Internal dimensions: w1200mmxD1000xH2100mm;

1.3Test temperature and allowable deviation temperature range: 35 ± 2 ℃;

1.4Salt spray settling amount: 1~2mL/80cm2 * h adjustable;

1.5Temperature fluctuation: ≤ 1 ℃;

1.6Temperature gradient: ≤ 2 ℃;

1.7Salt water concentration: 5% adjustable (mass ratio);

1.8Spray mode: continuous, adjustable cycle;

1.9Mist removal method: air compression;

1.10Test timing: 1~9999h, m, s adjustable;

1.11Test method: Neutral salt spray test.

2. material

2.1Shell: Imported high temperature, high humidity, and highly corrosive plastic board;

2.2Inner liner: Imported high temperature, high humidity, and highly corrosive plastic board;

2.3Box door: imported high temperature, high humidity, and highly corrosive plastic material; Set up a transparent observation window;

2.4Sample holder: Imported high temperature, high humidity, and highly corrosive plastic plate, with freely adjustable test angles of 15-30 degrees (angle with the vertical direction);

2.5Thermostatic water bath: PP impact resistant plate, equipped with an automatic water level controller;

2.6Saturated air tank: made of SUS304 stainless steel, with an electrolytic treatment on the surface and automatic water replenishment function;

2.7Reagent replenishment bucket: made of corrosion-resistant and high-temperature resistant plastic;

2.8Mist exhaust pipe: PVC hard plastic pipe.

3. spray system

3.1Spray device: tower type spray+non crystal glass nozzle, tower tube guidance, conical dispersion, and conical disperser to adjust spray amount;

3.2Salt collection: standard funnel and standard measuring cylinder;

3.3Pressure regulating valve: manually adjust the inlet pressure and spray pressure;

3.4Pressure gauge: indicates the pressure transmitted from the saturated air tank to the nozzle;

3.5Atomization treatment: equipped with filtering elements, the storage of atomized salt water is built-in hidden and has a large storage capacity, with salt water preheating function;

3.6Saturated air tank: filters, heats, humidifies air;

3.7Spray time: 1~9999h, m, s adjustable;

3.8Air compressor: discharge pressure greater than 8Kg, volumetric flow rate greater than 0.12 cubic meters/min;

4. temperature control

4.1Heater: Fully independent, nickel chromium alloy electric heater;

4.2Temperature sensor: PT100 platinum resistor;

4.3Controller: LED digital PID+SR microcomputer integrated controller.

5.other

5.1Operating environment temperature: 5~± 35 ℃;

5.2Operating environment humidity: ≤ 85% RH;

5.3Power supply (V): AC380 ± 10% V, 50 ± 0.5Hz;

5.4Attachment configuration: nozzle, measuring cup;

5.5Safety configuration: leakage protection, short circuit protection, compressor overpressure protection, compressor overload protection, overtemperature protection, water shortage protection, phase sequence/phase loss protection, end of test, overcurrent protection.

6. Execution and compliance with standards and testing methods

6.1GB/T10587Technical conditions for salt spray test chamber;

6.2GB/T2423.17Environmental testing for electrical and electronic products - Part 2: Test methods - Test Ka: Salt spray;