GC-9560-HDSpecialized for transformer oilThe chromatograph is a gas chromatograph used for determining the content of dissolved gas components in insulating oil for power use. The instrument adopts a three detector process and is equipped withTCDDetector and twoFIDDetector, one injection7Dissolve in insulating oil within minutes7Full analysis of gas component content, and detection if necessaryO2TheN2The content. among whichH2TheO2TheN2passTCDtesting;COTheCO2passFID1Detection, hydrocarbon gases(CH4TheC2H4TheC2H6TheC2H2)ByFID2Testing has overcome a lot of challengesCOTheCO2The impact on hydrocarbon gases, especially onC2H2The impact has greatly improved the detection sensitivity while shortening the detection time.
1、 Technical parameters:
1. Detection range and detection limit (unit µ L/L):
| component | Testing Scope |
| H2 | 2-10000 |
| CO | 2-50000 |
| CO2 | 2-50000 |
| CH4 | 0.1-2000 |
| C2H4 | 0.1-2000 |
| C2H6 | 0.1-2000 |
| C2H2 | 0.1-2000 |
2. Technical specifications of detector:
| Thermal conductivity detector(TCD) | Hydrogen flame ionization detector(FID) |
| Adopting a semi diffusive structure | Cylindrical collection electrode structure design, quartz nozzle, responsive |
| The power supply adopts constant current control mode | Automatic ignition; Stable time: Ten minutes |
| Sensitivity: ≥2500mV·ml/mg(Hexadecane)/Isooctane) | Detection limit: ≤8×10-12g/s(Hexadecane)/Isooctane) |
| Baseline noise: ≤20μV | Baseline noise: ≤2×10-13A |
| Baseline drift: ≤100μV/30min | Baseline drift: ≤2×10-12A/30min |
| line Sex: ≥104 | Linear: ≥106 |
3. Temperature control indicators:
| Temperature control location | Temperature control range | temperature control accuracy |
| pillar box | At room temperature8℃~400℃ | ±0.1℃ |
| Hydrogen flame(FID)Detector | At room temperature8℃~400℃ | ±0.1℃ |
| hot Guide(TCD)Detector | At room temperature8℃~400℃ | ±0.1℃ |
| NiCatalytic converter | At room temperature8℃~400℃ | ±0.1℃ |
2、 Execution standards:
GB/T 17623-1998 "Determination of dissolved gas components in insulating oil by gas chromatography"
GB/T 7252-2001 "Guidelines for Analysis and Determination of Dissolved Gases in Transformer Oil"
threeComputer anti control function:
After the analysis is completed, multiple fault diagnosis methods are provided, including automatic over limit prompts, three ratio diagnostics that comply with national standards, TD diagrams, component concentration diagrams, David triangles, etc.