The GC-9560-FT hydrogen flame thermal conductivity gas chromatograph uses dual injection and one instrument to achieve full analysis of inorganic impurities and hydrocarbon impurities in industrial grade gases. The detector adopts high-sensitivity hydrogen flame detector and thermal conductivity pool detector, and is equipped with an efficient conversion furnace to ensure high detection limit:
1、 Technical indicators
1. Detection limit: (using ultra pure hydrogen (99.9999%) as carrier gas)
| O2 | N2 | CO | CH4 | CO2 |
| 5 | 5 | 10 | 10 | 10 |
2. Overall noise ≤ 5UV
Baseline drift:<100UV/30min
1ml injection volume, peak height of 1ppmN2>15UV
3ml injection volume, peak height of 1ppmN2>50UV
3. System configuration:
GC-9560 gas chromatograph 1 set
1 set of FID detector
1 set of TCD detector
1 set of conversion furnace
1 set of chromatography workstation
2、 Technical features:
1. The pipeline adopts imported materials and is hard sealed, ensuring no dead volume, adsorption or diffusion in the system to ensure the sensitivity and accuracy of thermal conductivity;
2. The control of the air path adopts a rear mounted control to avoid air infiltration from the header into the analyzed sample;
3. The sample can achieve baseline separation;
4. The tail of the valve injection system is equipped with a rotor flowmeter and a needle valve to control the injection flow rate and ensure the repeatability of the injection volume.