In order to meet the needs of power system analysts who need to carry instruments to the fault site far away from the laboratory for transformer oil analysis, our Shanghai Huaai Chromatography Analysis Technology Co., Ltd. has newly designed the GC-9760A portable chromatograph for transformer oil. The instrument integrates all components such as computers and gas sources, and the number of components in the entire system is compressed into 2, achieving true portability.
This instrument adopts the classic three detector process, which can complete the dissolution of H in insulating oil with one injection2、CO、CO2、CH4、C2H4、C2H6、C2H2Wait for a full analysis of the content of 7 gas components, and if necessary, O can also be detected2、N2The content.
Execution standard:
GB/T 17623-2017 "Determination of dissolved gas components in insulating oil by gas chromatography"
GB/T 7252-2001 "Guidelines for Analysis and Determination of Dissolved Gases in Transformer Oil"
DL/T 703-2015 "Gas Chromatography Determination of Gas Content in Insulating Oil"
Technical Specifications:
1Detection range and detection limit (unit µ L/L):
| component | Testing Scope |
| H2 | 2-10000 |
| CO | 2-50000 |
| CO2 | 2-50000 |
| CH4 | 0.1-2000 |
| C2H4 | 0.1-2000 |
| C2H6 | 0.1-2000 |
| C2H2 | 0.1-2000 |
2. Technical specifications of detector:
| Thermal conductivity detector(TCD) | Hydrogen flame ionization detector(FID) |
| Adopting a semi diffusive structure | Cylindrical collection electrode structure design, quartz nozzle, responsive |
| The power supply adopts constant current control mode | Automatic ignition; Stable time: Ten minutes |
| Sensitivity: ≥2500mV·ml/mg(Hexadecane)/Isooctane) | Detection limit: ≤8×10-12g/s(Hexadecane)/Isooctane) |
| Baseline noise: ≤20μV | Baseline noise: ≤2×10-13A |
| Baseline drift: ≤100μV/30min | Baseline drift: ≤2×10-12A/30min |
| line Sex: ≥104 | Linear: ≥106 |
3. Temperature control indicators:
| Temperature control location | Temperature control range | temperature control accuracy |
| Column box | At room temperature8℃~400℃ | ±0.1℃ |
| Hydrogen flame(FID)Detector | At room temperature8℃~400℃ | ±0.1℃ |
| Thermal conductivity(TCD)Detector | At room temperature8℃~400℃ | ±0.1℃ |
| NiCatalytic converter | At room temperature8℃~400℃ | ±0.1℃ |
4. Other parameters:
Dimensions: 380 * 450 * 180mm
Weight:<15kg
Power supply: AC220V ± 10 50Hz
SZD-5 Small Fully Automatic Oscillator
Technical Features
1. Computer reverse control technology, all operations can be controlled through a chromatography workstation
2. Can oscillate two oil samples simultaneously
3. Upper door design for easy access to needle tubes
technical indicators
1. Oscillation frequency: 275 ± 5 times/minute
2. Amplitude: 36mm
3. Dimensions: 255mm × 380mm × 200mm
4. Weight: 8Kg
5. Power supply: AC220V 50Hz
6. Power: 200W