The Mingcheng Laser Smoke and Dust Direct Reader MC-5YC achieves the detection of dust concentration in areas such as air conditioning exhaust outlets, flues, wind tunnels, and deep wells by adding a flue sampling rod. In the standard configuration, the total length of the sampling rod and hose is 1.6m, and the sampling rod can be customized to be extended according to customer needs, with a length not exceeding 3m. This instrument uses a powerful suction pump, which is not affected by environmental pressure, and is equipped with a 300 degree high-temperature resistant sampling rod, which still performs well even in high temperature environments. This product is a flue dust detector that is easy to carry, operate, and measures quickly and accurately.
Mingcheng Laser Flue Gas Direct ReaderMC-5YC (Highly detectable 300 degree smoke and dust)
The laser soot analyzer achieves the detection of dust concentration in areas such as air conditioning exhaust outlets, flues, wind tunnels, and deep wells by adding a flue sampling rod. In the standard configuration, the total length of the sampling rod and hose is 1.6m, and the sampling rod can be customized to be extended according to customer needs, with a length not exceeding 3m. This instrument uses a powerful suction pump, which is not affected by environmental pressure, and is equipped with a 300 degree high-temperature resistant sampling rod, which still performs well even in high temperature environments. This product is a flue dust detector that is easy to carry, operate, and measures quickly and accurately.
Mingcheng Laser Flue Gas Direct ReaderMC-5YC Technical parameters for measuring up to 300 degrees of smoke and dust
1. Configure a 40mm filter membrane sampler;
2. Featuring replaceable particle cutters PM10, PM5, PM2.5, and TSP for selection;
3. Direct reading of dust mass concentration (mg/m ³), results can be obtained in 1 minute, or can be set as needed;
4. Large screen LCD display with Chinese character menu prompts;
5. Detection sensitivity: 0.001mg/m ³;
6. Repetitive error: ± 2%
7. Indication error: ± 10%
8. Measurement range: 0.001-10 mg/m ³.
9. Measurement time: The standard time is 1 minute, with 0.1 minutes and manual mode (sampling time can be set arbitrarily).
10. It has public place monitoring mode, atmospheric environment monitoring mode, and labor hygiene mode. Time weighted average (TWA) and short-term exposure allowance concentration (STEL) can be calculated.
11. Storage: It can cyclically store 999 sets of data.
12. Timed sampling (automatic continuous measurement mode): The measurement duration can be set to (1~9999) seconds, the interval time can be set to (0~9999) seconds, and the sampling frequency can be set to (1~9999 times).
13. Equipped with an alarm function for excessive dust concentration, the alarm concentration threshold can be set by oneself; The measurement duration (1~9999) seconds can be set by oneself.
14. Output interface:
(1) PC communication serial interface: RS232;
(2) Micro printer output interface;
(3) Analog output interface: (0-1) V;
(4) Digital output interface: level signal.
15. Power supply: Lithium ion rechargeable battery pack, capable of continuous use for 8 hours; Attached is a 220VAC/12VDC power adapter.
16. Equipped with humidity correction function, the data is more accurate
17. Equipped with a 300 degree high-temperature resistant sampling rod and hose, with a total length of 1.6m, which can be extended by 3m.
18. Weight: 2.4kg