When measuring, use a single lens reflex method with no deviation between the actual measurement field and the measurement field inside the viewfinder. Only measure within the measurement range,
It can avoid the influence of light sources outside the measurement range on the optical system of non expanded tube bodies.
The measurement angle of LS-100 is 1 °, while LS-110 is 1/3 °, so it can measure the brightness of small parts. Additionally, by using close-up shots (sold separately),
LS-100 can measure a diameter of 1.3mm,LS-110 can measure the brightness of small areas within 0.4mm.
As an arbitrary calibration mode, users can input any calibration reference that can be recalibrated on their own standard brightness,
It is also possible to use multiple luminance meters in the * state to implement precise luminance management by inputting a color modification coefficient that modifies the indication difference to a specific light source.
| form | Point type DSLR digital brightness meter |
| measure angle | 1° |
| optical system | f=85mmF2.8 |
| Viewfinder field of view | 9 ° (with viewfinder view adjustment function) |
|
measure distance(measure area)
| ∞~1,014mm(∞~φ14.4mm) |
| Starting from the benchmark position mark of the self body | When using close-up shots 53 1210~623mm(φ18.7~φ8.0mm) 35 615~447mm(φ8.7~φ5.2mm) 22 368~323mm(φ4.3~φ3.2) 10 205~203mm(φ1.5~φ1.3mm) |
| Photo detector | Silicon photocell |
| Spectral sensitivity | Deviation fs from standard spectral luminous efficiency (V λ): within 8% |
| Response time | FAST: Display time is 0.8-1.0 seconds, sampling time is 0.1 seconds SLOW: Display time is about 1.4-1.6 seconds, sampling time is 0.4 seconds |
|
Luminance unit※
| cd/m2
|
| Measure brightness range | FAST: 0.001~299900cd/m2 SLOW: 0.001~49990cd/m2 |
| measurement accuracy | 1,000cd/m2Above: ± 2% ± 1 digit of the indicated value 0.001~0.999cd/m2± 2% ± 2digit of the indicated value (under conditions of A light source and 20-30 ℃) |
| repeatability accuracy | 1,000cd/m2Above: ± 0.2% ± 1 digit of the indicated value 0.001~0.999cd/m2± 0.2% ± 0.2 digit of the indicated value (A light source) |
| temperature error | ± 3% ± 1 digit of the indicated value (0-40 ℃, 20 ℃ standard) |
| calibration mode | Konica Minolta/Any Benchmark Conversion Enter any calibration benchmark: 1 channel, input numerical value Input color correction factor: 1 channel, input value (0.001-9.999) |
| Brightness ratio mode | 1 channel, input measurement value or numerical value |
| Measurement function | Measure instantaneous brightness value, peak brightness value, and brightness ratio (%) |
| express | External display: LCD has 4 significant digits Display in viewfinder: LCD with LED illumination, effective number of digits is 4 |
| Digital signal input and output | Using RS-232C as the standard Data transmission speed 4800BPS |
| external operation | Can be controlled from the digital output terminal |
| power supply | The power supply is one 9V stacked dry battery (JIS 6F22) that can be provided from the data printer DP-10 |
| power consumption | When the display light in the viewfinder is on, the average battery level is 16mA (with the measurement button in the ON state) When the display light in the viewfinder goes out |
| Usage environment | Temperature and humidity: 0-40 ℃, relative humidity below 85% (at 35 ℃)/no condensation allowed, Pollution level: 2, equipment type: II |
| Storage temperature and humidity range | -20-55 ℃, relative humidity below 85% (at 35 ℃)/no condensation allowed |
| Size and weight | 79 (width) x 208 (height) x 150 (depth) mm 850g (excluding battery) |
| Standard Accessories | Lens cover, eyepiece ND filter, eyepiece cover, 9V laminated dry battery (JIS 6F22), box |