CS-200
color brightness meter
The CS-200 colorimeter can measure the brightness and chromaticity of various luminescent products such as large-sized plasma displays, small-sized LCDs and LEDs, outdoor screens, high-voltage lamps, and instrument panels.
The new product launched in the spring of 2006 will be equipped with a new automatic mode that can measure low brightness with high precision.
This newly launched automatic measurement mode can automatically select the measurement speed based on the brightness of the measured object.
This function is particularly suitable for measuring low brightness objects and can effectively avoid the phenomenon of poor measurement accuracy caused by measurement speed.
By using this automatic measurement mode, the measurement speed can be automatically selected to maximize the measurement capabilities of the CS-200 and easily obtain high-precision measurement results.
type
Measure brightness, chromaticity, and related colors for research and inspection purposes.
main purpose
Various optical devices such as LCD, PDP, organic EL, and FED, as well as light sources such as LED and lamps.
Color Brightness Meter - Measuring the color visible to the naked eye through sensors
Color Brightness Meter CS-200
feature
The use of spectral function can achieve high-precision measurement of brightness, chromaticity, and correlated color temperature for measuring objects such as lamps and displays.
purpose
In order to complete R&D or inspection of organic EL and FED for LEDs, LCDs, PDPs, lamps, and next-generation displays, this product can be used for measuring brightness, chromaticity, and related colors.
Product Introduction
Accuracy approximate spectrophotometer for light source incident colorimeter
The CS-200 can achieve high-precision brightness and chromaticity measurements comparable to spectroscope instruments, while retaining the simplicity, ease of use, and affordability of tristimulus instruments.
The CS-200 is equipped with 40 sensors and calculates using a spectral sensitivity characteristic curve (isochromatic function) that matches the sensitivity of the naked eye. Through this developed spectral fitting method, more accurate tristimulus values (XYZ of red, green, and blue) can be obtained than traditional tristimulus colorimeters, achieving higher precision.
It can measure a wide range of brightness levels (from low brightness of 0.01 cd/m2 to high brightness of 20000000 cd/m2, measured at an angle of 0.1 °).
Three measurement angles can be selected: 1 °, 0.2 °, and 0.1 °. The measurement angle can be switched according to the target being measured (from display devices or other objects that require measuring a large area to objects that require measuring a small area such as car audio control panels, car instrument panels, and small LCDs).
Application field
All types of light sources can be measured using CS-200, such as signal lights, traffic lights, airport lighting, lights, etc LED、 Almost all objects that emit light, such as cathode ray tubes, LCDs, PDPs, etc. Due to the development of information technology business, the demand for diversified media in the market is gradually increasing. In addition, with the launch of comprehensive digital broadcasting services, the research and development of various display devices centered on plasma display screens and LCDs has accelerated. The innovation of LED technology has brought significant breakthroughs to the development of LED products, including traffic lights, backlighting for reflective LCDs, and large-sized outdoor display screens.
background knowledge
There are two ways to determine the brightness and chromaticity values of a light source.
The three stimulus type instrument uses three sensors (red, green, blue) to measure the light source, and the sensitivity of these sensors is similar to the sensitivity of the naked eye to light. The filter used to achieve this sensitivity has limited accuracy, so the three stimulus instrument cannot provide accurate readings for all types of light sources. If a certain spectral distribution (similar to the spectral distribution of the light source used for calibrating instruments) is used to measure the light source, accuracy can be achieved.
The CS-200 uses this second method, which measures the spectral emissivity of a light source and calculates it using a spectral sensitivity characteristic curve (isochromatic function) that matches the sensitivity of the naked eye. By using this method, it is possible to avoid filter mismatch and obtain more accurate measurement results.
Product Features
New automatic mode
The new automatic mode can adjust the measurement speed based on the brightness of the object being measured. It may not be possible to obtain the necessary accuracy when measuring at certain measurement speeds, especially in low brightness situations. The new automatic mode will automatically select the measurement speed of * to provide the performance of CS-200 *, making it easier to achieve higher measurement accuracy.
main purpose
CS-200 can be used to measure the brightness and chromaticity of various optical devices such as LCD, PDP, organic EL, and FED, as well as light sources such as LED and lamps.
Main features
Wide measurement range, covering both low and high brightness.
Performance comparable to multiple spectroradiometers
The body is compact and lightweight. It can also be powered by batteries.
Measurement angle optional
Provide data management software CS-S10w standard version and optional professional version.
Other functions
The measurement and display devices can be synchronized by inputting frequency values. Measurement speed options (automatic, limited automatic, manual, ultra fast, fast, slow, and ultra slow) with backlit large LCD display and USB 1.1 communication
Data storage: 101 measured values (with 9-letter IDs) and 20 reference values
User correction: 20 channels
Product Information
Product Overview
This product adopts a newly developed spectral fitting method, which can be applied to various displays LED、 High precision measurement of brightness, chromaticity, and correlated color temperature of lighting. This product is compact and lightweight, with a battery powered mode that is easy to operate. It can measure from low brightness to high brightness with just one device, making it economical and practical.
main purpose
This product can accurately measure the brightness, chromaticity, and correlated color temperature of various light sources, from LCD and PDP to organic EL and FED for next-generation displays, and then to optical devices such as LED and lighting.
Main features
This product adopts spectral fitting method, which can accurately measure the monochromatic brightness and chromaticity of various displays, and the measurement accuracy is comparable to that of a spectral radiance analyzer.
Synchronous measurement can be performed on display devices.
From a low brightness of 0.01cd/m2 to a high brightness of 20000000cd/m2 (measured at an angle of 0.1 °), the measurement range is wide.
Convertible measurement angle.
The measurement device can be controlled through the data management software CS-A10w installed on the computer to perform various data management, analysis, and evaluation of measurement data.
The portable design is compact and lightweight, with a simple 3-type 4-cell battery power supply mode. In addition, an AC connector is also provided for use.
Specification (sample 1)
Main specifications
|
model | CS-200 |
| measurement range | 0.01-200000 cd/m2 (measurement angle 1 °) |
| accuracy | 150 cd/m2 (for light source A): Lv ± 2% ± 1 bit; xy ±0.002 |
| repetitiveness | 0.01-1 cd/m2 (for light source A): Lv ± 0.01 cd/m2 ± bit; ---(2 σ/automatic) |
| measure time | Automatically set between 1 second and 60 seconds (automatic); Automatically set to 1 second or 3 seconds (limited automatic) |
| measurement method | Spectral method, grating+linear photodiode array |
| measure angle | 1 °, 0.2 °, 0.1 ° (switchable) |
| Zui small measurement area | 0.5 mm |
| Zui small measurement distance | 296 mm (distance to the front edge of the metal lens barrel) |
| Observation conditions | 2/10 degrees |
| Color space | Lv x y、Lv u' v'、Lv TΔuv、XYZ、 dominant wavelength |
| Measurement synchronization setting range | Vertical synchronization frequency: 40.00~200.00Hz |
| interface | USB 1.1 |
| power supply | AC adapter or 4 AA batteries |
| Battery lifespan | Approximately 3 hours (continuous measurement/fast mode/AA type alkaline battery) |
| size | 95 mm (W) × 127 mm (H) × 330 mm (L) |
| weight | 1.8 kg (excluding battery) |
| Temperature/humidity range for use | 0-40 ° C, relative humidity below 85% (35 ° C), no condensation allowed |
| Storage temperature/humidity range | 0-45 ° C, relative humidity below 85% (35 ° C), no condensation allowed |
#1 23 ° C ± 2C Lv=0.01-10 cd/m2, slow mode, average of 30 measurements
Lv=10 cd/m2 and above, slow mode, average of 10 measurements
#When the measurement angle is 0.2 °, the received light amount is approximately 1/25 of that at 1 °.
Therefore, the brightness targeted by the same repeatability is 25 times lower than that at 1 °.
When the measurement angle is 0.1 °, the received light amount is approximately 1/100 of that at 1 °.
Therefore, the brightness targeted by the same repeatability is 100 times lower than that at 1 °.
*Specifications are subject to change without prior notice.
Specification (sample 2)
|
form | Color illuminance meter | |
| Spectral sensitivity | The CIE2 ° field of view standard observer approximates the CIE 1931 isochromatic function | |
| Photo detector | Silicon photocell | |
| Measurement function | Three stimulus values | XYZ |
| Chromaticity Coordinates | Ev xy、Ev u'v' | |
| correlated color temperature | Ev、Tcp、 △uv | |
| Color difference display | Δ(XYZ)、Δ(Ev xy)、Δ(Ev u' v')、ΔEvΔu'v' | |
| Other functions | Calibration function, data retention, multi-point measurement (up to 30 light receiving units) | |
| measurement range | 0.1 to 99990 lx (chromaticity representation above 5 lx) | |
| measurement range | Automatic (4 ranges) | |
| Illumination level | JIS C 1609-1993 based on regular type A * | |
| measurement accuracy | Ev: ± 2% ± 1digit of the indicated value ± 2% ± 1digit of the value | |
| repetitiveness | Xy: ± 0.0005 (standard light source A) | |
| temperature characteristics | Ev: ± 3% ± 1 digit of the indicated value | |
| Humidity characteristics | Ev: ± 3% ± 1 digit of the indicated value | |
| measurement period | 2 times/second (continuous) | |
| Data signal output input | Using RS-232C as the standard | |
| express | LCD 4-digit, equipped with rear lights (automatically lit) | |
| Usage environment | temperature and humidity | -10-40 ℃, relative humidity below 85% (at 35 ℃) |
| pollution level | 2 | |
| Equipment Type | II | |
| Storage temperature and humidity range | -20-55 ℃, relative humidity below 85% (at 35 ℃)/no condensation allowed | |
| power supply | Single 3-type alkaline dry battery or AC adapter | |
| battery life | Approximately 72 hours (when using alkaline batteries for continuous measurement) | |
| Size and quality | 69 (width) x 174 (height) x 35 (depth) 215g (excluding battery) | |
| Standard Accessories | Soft box, cover, fastening clamp, battery | |
| Sell accessories separately | Light receiving unit, body connector T-A20, light receiving unit adapter T-A21, AC adapter AC-A10, data management software CL-S1w, printing cable T-A12, connecting cable T-A11 (for PC), protective cover, hard box | |