| Overview Glass rotor flowmeter is mainly used in various departments such as chemical, petroleum, light industry, medicine, fertilizer, synthetic fiber, food crystal, dye, environmental protection, and scientific research to measure the flow rate of single-phase non arterial (liquid or gas) fluids. Corrosion resistant flow meters are mainly used for flow detection of corrosive liquids and gas media, such as strong acids (except hydrofluoric acid), strong bases, oxidants, strong oxidizing acids, organic solvents, and other corrosive gas or liquid media feature 1. Low pressure loss 2. Reliable performance, convenient reading, and direct display 3. Simple structure, easy to install and use 4. Cheap price
Principle and Structure The main measuring components of the flowmeter are a vertically installed lower small upper large conical glass tube and a float that can move up and down inside the tube.When the fluid flows from belowWhen flowing through a conical glass tube, a pressure difference is generated between the upper and lower parts of the float, and the float rises under this pressure difference. When the upward force equals the buoyancy and viscous lift acting on the float and the weight of the float, the float is in an equilibrium position. Therefore, there is a certain proportional relationship between the fluid flow rate flowing through the flowmeter and the rising height of the float, that is, the flow area of the flowmeter. The position height of the float can be used as a measure of flow rate. Model specifications and technical indicators 普通型
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Note: The above specifications (DN2~10mm) of flow meters are all side inlet and side outlet, connected by hoses or threads, with needle type flow regulating valves on the lower base. For DN ≥ 15mm, they are flange type, installed vertically, without regulating valves.
|
Note: The liquid flow rate of the glass rotor flowmeter is calibrated with a scale of 20 ℃ clean water, and the gas flow rate is calibrated with a scale of 20 ℃ and 101325Pa air; According to user needs, special flow meters can be made, with a DN100 diameter water flow of up to 120m3/h, The airflow can reach 3500m3/h. Special requests can also be made Calibration of medium density correction and calibration of actual pressure correction according to specific operations. anti-corrosive
Note: ※ Corrosion resistant models with nominal diameter DN ≤ 10mm need to be customized and generally do not come with needle type regulating valves; According to user requirements, we can produce corrosion-resistant flow meters with a higher level of precision. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||

Material of the contact part between the flowmeter and the measured medium
普通型
| LZB-2 LZB-3 LZB-4 LZB-6 LZB-10
| |||
| Base and valve needle | stop file | float | sealing packing |
| Brass chrome plating | Polytetrafluoroethylene (PTEE) | Agate ballor (304) stainless steel | Acid alkali resistant rubber (IV-1) |
| LZB-15 LZB-25 LZB-40
| |||
| flange | stop file | Float and guide rod | sealing packing |
| Cast iron interior Lining acid alkali resistant rubber (IV-1) | Polytetrafluoroethylene (PTEE) | (304) stainless steel | Acid alkali resistant rubber (IV-1) |
| LZB-50 LZB-80 LZB-100
| |||
| flange | stop file | Float and guide rod | sealing packing |
| Cast iron inner coating Amino baking paint | Cast iron inner coating Amino baking paint | (304) stainless steel | Acid alkali resistant rubber (IV-1) |
|
Note: When used in conjunction, our company can make changes to the appearance and structural dimensions of the flowmeter according to user requirements, with special designs.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||