1Full temperature shaker (constant temperature oscillator)Product Introduction:
XS-CFull temperature shaker/The oscillator, also known as the air constant temperature shaker, is a biochemical instrument that combines a temperature controlled incubator and an oscillator. It is a laboratory equipment used for precision cultivation and preparation in research, education, and production departments such as plants, biology, microbiology, genetics, viruses, environmental protection, and medicine. The main features of the constant temperature shaker are: ① Accurate temperature control and digital display. ② Set up oxygen replenishment holes and ensure sufficient oxygen replenishment in the constant temperature working chamber. ③ Equipped with mechanical timing and intelligent timing for temperature control. ④ The spring bottle rack is particularly suitable for the cultivation and preparation of biological samples for various comparative experiments. ⑤ Infinite speed regulation ensures smooth operation. Easy to operate and safe.
2Full temperature shaker (constant temperature oscillator)Technical Indicators:
Speed range: Start -300r/min
Temperature control range: 5-50 ℃ (maximum up to 60 ℃)
Temperature control fluctuation: ± 0.5 ℃
Uniform temperature control: ± 0.1 ℃
Temperature control accuracy: 0.1 ℃
Amplitude: 30mm
Heating power: 800W
Bottled volume: 1000ml × 30
Oscillatory timing: 0-120min or normally open
Temperature control timing: 0-99 hours
Power supply: 220V 50Hz
Instructions for Use:
1. Using at medium speed within the speed range can extend the lifespan of the instrument.
2. The full temperature oscillator should be placed on a flat surface, and the environment should be kept clean, tidy, ventilated, and dry.
3. Before using the instrument, place the speed control knob in the minimum position and turn off the "oscillation switch".
4. Packaging test bottles: ① Pay attention to uniform distribution. ② Pay attention to sealing the bottle mouth. ③ Prevent condensed water droplets from dripping into the test bottle.
5. Connect the external power supply, place the power switch in the "on" position, and the indicator light will turn on.
6. Temperature control:
① Turn on the power and the self-test will end in about one minute. If "000" is displayed, it indicates that the sensor is open or the input signal exceeds the measurement range.
② Press the SET key to set the temperature. Press the SET key until the data in the lower row of the convergence tube flashes, indicating that the instrument has entered the temperature setting state. Press the △ key to increase the set value, press the ▽ key to decrease the set value, and then press the SET key again to return to the normal working state. The temperature setting is complete.
③ When confirming that the value displayed on the instrument is not the correct measurement value, the displayed value can be corrected. Press the SET key for 3 seconds to enter the inner menu of the instrument panel. After selecting parameter SC, use the △ or ▽ keys to modify this parameter. The range of correction for sensor errors is 50 to
-50. After the correction is completed, press the SET key for 3 seconds to exit. The correction value of the instrument is 0 when it leaves the factory. When using it, it is necessary to prevent the instrument that displays correctly from being corrected to incorrect.
④ The self-tuning function of the instrument: This product can adapt to the vast majority of control systems, and can only start self-tuning when the control effect is particularly unsatisfactory. Press the SET button for 3 seconds to enter the inner menu of the instrument, select the parameter At, and use the △ or ▽ button to start self-tuning. At this time, the At indicator light starts flashing, and the instrument enters the automatic tuning parameter working state. The instrument controls the heating system to fluctuate twice near the set point. When the At light goes out, self-tuning ends, and the new PID parameters are locked in the chip.
7. Turn on the oscillation switch, the indicator light will turn on, slowly adjust the speed control knob to the desired speed.
8. After completing the work, set the temperature control button and speed control knob to the minimum value and cut off the power.
5、 Maintenance and upkeep:
1. Proper use and maintenance of the instrument can prolong its service life by keeping it in good working condition.
2. The instrument should have good grounding to prevent accidents. It is strictly prohibited for various solutions to enter the host room to avoid damaging the host.
3. The bearings of the transmission part have been filled with an appropriate amount of lubricating grease before leaving the factory. During continuous operation of the full temperature oscillator, lubricating grease should be added every six months, with a filling amount of about one-third of the bearing space.