The FKC-I type airborne dust and bacteria sampler is an efficient porous inhalation dust and bacteria sampler. Newly upgraded touch operating system with fast response,
The human-computer interaction interface is user-friendly. The instrument is designed based on the theory of constant velocity sampling, with direct sampling. The wind speed at the collection head is basically the same as that in the clean room, which can provide more accurate results
Reflect the microbial concentration in the cleanroom. During sampling, the air carrying dust bacteria passes through the micropores of the sampling head at high speed and evenly impacts the surface of the agar in the culture dish;
These living microorganisms obtain nutrients evenly and thoroughly on the agar surface, and during the cultivation process, they undergo rapid dynamic rehydration and high-speed growth, resulting in more
Get the results quickly.
2、 Usage environment
1. Temperature: 10-35 ℃
2. Relative humidity: 10-90% RH
3. Atmospheric pressure: 80-110kPa
4. Dust concentration: 100000000 particles/m ³, 0.5 µ m or 0.2mg/m ³
3、 Main features and parameters:
1. Sampling head mouth: The sampling head has countless micropores, allowing microorganisms to be evenly distributed on the surface of the agar, reducing the overlap of dust and bacteria,
Reduced microbial counting errors.
CTP 2.8 Interactive System
3. Sampling flow rate: 100L/min.
4. Sampling head velocity: 0.38m/s, which is basically the same as the wind speed in the clean room (constant velocity sampling design).
5. Sampling volume: The sampling volume can be set arbitrarily from 0.001-9.999m ³.
6. Culture dish: Use standard universal culture dish with a diameter of 90 * 15
7. Power supply: DC16.8V, can work continuously for 7 hours after being charged.
8. Dimensions: Φ 120 * 325.
9. Weight: 3.0kg