Overview
The method of using gas chromatography for analysis in the electronic moisture analyzer MOC63u is called gas chromatography. The principle of gas chromatography is that after the sample solution is injected, it first enters the vaporization chamber, and then enters the chromatographic column under the transport of the carrier gas (usually nitrogen or helium gas). Different components are separated in the chromatographic column, and then flow out of the chromatographic column in sequence, which is detected by the detector to obtain their content.
Product Details

Electronic moisture analyzer MOC63uEasy to operate - automatic start mode

Easy operation mode and keyboard design, automatic heating mode, saving measurement time

Large weighing tray

The sample disk size is: diameter 95mm

Adopting a backlit LCD display
compact design

Among electronic moisture meters, MOC63u is the most compact product, with a length of only 202 (mm).

Halogen lamp heating

Halogen lamp heating provides you with fast and accurate measurements.

Electronic moisture analyzer MOC63uData Management - Windows Direct Window

The measurement conditions and data can be stored in MOC63u. There is a DATA IO interface for connecting to a printer, and RE232C and USB interfaces for connecting to a PC. Data can be directly transferred to EXCEL and other Windows applications.

repair

Cleaning and replacing halogen lamps is still very simple.


MOC63u measurement method
Please choose the correct measurement method

Stop method

Automatic stop mode
When the change in moisture content within 30 seconds reaches below the calibrated automatic stop condition, the heating will automatically stop and the measurement will end.

Timed stop mode
When the preset time is reached, the heating will automatically stop and the measurement will end.

Interval drying method

Quick drying method
At the beginning of the measurement, heat up at a high temperature first, and then heat up again at the calibrated temperature to shorten the heating time.

Slow drying method
According to the usual measurement, slowly increase the drying temperature and heat the sample.

Step by step drying method
In dry conditions, heating conditions can be set in multiple steps. This is a very good measurement method for samples containing a large amount of moisture.

Startup method
When the upper cover is closed, it will immediately heat up for measurement. In repeated measurements, this can be very time-saving.


range MAX 60g
MIN 0.02g
Zui small display value 0.001g
0.01%/0.1% (optional)
repetitiveness 0.15%(2g)
0.05%(5g)
0.02%(10g)
heat source Halogen lamp heating
power 400W
Temperature setting range 50-200 ℃ (step distance 1 ℃)
display screen LCD backlit display
Scale size Diameter 95mm
host size 202×336×157
host weight 4kg
Working temperature and humidity range 5-40 ℃, below 85% RH
measurement method Standard mode, fast drying, slow drying, segmented drying
Measurement time 1-120 minutes (up to 12 hours)
interface RS232C (9-pin), DATA I/O, USB interface
Measurement condition storage 10 of them
data storage 100 sets
Temperature calibration component Optional accessories