Product principle: Conductivity: a physical concept, also known as conductivity

Product Details

Product Principle
Conductivity: a physical concept, also known as electrical conductivity. The product of this quantity and the electric field strength E in the medium is equal to the conduction current density J. For isotropic media, conductivity is a scalar; For anisotropic media, conductivity is a tensor. In ecology, conductivity is the ability of a solution to conduct electrical current, represented by numbers. Unit expressed in Siemens per meter (S/m)
Conductivity is a parameter used to describe the ease of charge flow in a substance. The standard unit for conductivity σ is Siemens/meter (abbreviated as S/m), which is the reciprocal of resistivity ρ.
Conversion of conductivity units: The standard unit for single conductivity is Siemens per meter (S/m), where S is Siemens, 1S=1000mS=1000000 μ S, 10000us/cm=1 S/m μ is a Greek letter representing 10 to the negative 6th power, pronounced as "micro"; S is the international unit of electrical conductivity, pronounced as' Siemens'; Cm is the unit of length in centimeters, 1cm=10-2m. μ s/cm is pronounced as micrometer per centimeter and is a unit of electrical conductivity. Conversion: 1 US/cm=10 ^ -4 S/m
Application field
☛ Waterworks/seawater desalination
In terms of wastewater treatment and sewage testing
Mining and smelting, papermaking, and pulp concentration
Environmental protection industry, desulfurization and denitrification, wastewater evaporation
Medical and health applications, food and beverage processing industry
Experimental or testing equipment, thermal power generation, petrochemicals, textile drawing and other industries
Product Features
There are multiple electrode constants to choose from
K=0.01 (used for measuring ultrapure water and pure water), K=0.1 (used for measuring drinking water and tap water), K=1 (used for measuring raw water and conventional chemical water), K=10 (used for measuring sewage)
Technical Specifications
Measuring range 0~20µS·cm1(K=0.01 代号 K1); 20~200µS·cm1(K=0.1 代号 K2); 200µS·cm1~10mS·cm1(K=1.0 code K3); 10~20.0mS·cm1(K=10.0 code K4)
Resolution value: Conductivity: 0.01 µ S · cm1;0.01 mS·cm1; Temperature: 0.1 ℃;
Comprehensive accuracy: Conductivity: 1%; Electrical resistivity 1.5%
Ambient temperature -10~60℃
relative humidity Not exceeding 90%
Body material: 316L stainless steel material
Voltage endurance range: ≤6MPA
Connection method: 1/2NPT pipe thread connection (other methods with adapter) can be provided according to user needs, including flow type, pipeline type, sinking type, flange type (default thread)
Outgoing method: Direct outgoing line (A); Y-shaped interface (B); OT interface (C) (default direct output)
Temperature compensation: NTC 10K thermistor (default)
Special reminder: There is no strong magnetic field interference around except for the Earth's magnetic field
Ordering Guide
model Electrode constant install thread qualification method none none Other requirements
MX-DD-01 K3 1/2NPT Direct outgoing line none none ---
Ordering example: MX-DD-01-K3
schematic diagram
process steps
1. Coarse filtration: Real time monitoring of pH value, pressure, etc. inside the tank is required during the treatment of irrigation
2. Softening: This is the second step of the process, which involves removing excess Ga and Mg ions and requires a conductivity meter
3. Filtering: requires instruments such as pressure and flow rate
4. Disinfection and reverse osmosis: This process requires monitoring of temperature, conductivity, pressure, and integrated display
5. Membrane degassing: requires the use of pressure, temperature, differential pressure and other instruments
6. Disinfection and Electrodialysis (EDI): This process is an important step in producing pure water and requires monitoring of products such as pH, conductivity, and temperature.
7. Storage tank: liquid level needs to be monitored and recorded, requiring a liquid level gauge and paperless recorder
Special reminder