Carbon hydrogen analyzer

NegotiableUpdate on 03/13
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The KY-300 microcomputer hydrocarbon analyzer instrument uses Coulomb method to measure hydrogen and gravimetric method to measure carbon. The computer operates and processes the data, and the accuracy and precision meet the standards. This instrument is used to determine the content of organic and inorganic carbon and hydrogen in coal and other solid materials. It is one of the instruments used by coal mines, quality inspection, power plants, geological exploration, metallurgy and other departments to detect the content of carbon and hydrogen. It has a wide range of applications in coal, power, chemical, refining, commercial inspection and other units.
Product Details

KY-1Microcomputer based hydrocarbon analyzer

The instrument uses Coulomb method to measure hydrogen and gravimetric method to measure carbon, and computer operation and data processing meet the standards for accuracy and precision. This instrument is used to determine the content of organic and inorganic carbon and hydrogen in coal and other solid materials. It is one of the instruments used by coal mines, quality inspection, power plants, geological exploration, metallurgy and other departments to detect the content of carbon and hydrogen. It has a wide range of applications in coal, power, chemical, refining, commercial inspection and other units.

Main features of the instrument:

1

Computer control, simple and accurate, with good repeatability, suitable for the determination of carbon and hydrogen content in coal and other organic compounds.

2

Users can set the instrument as a hydrocarbon analyzer or a hydrogen analyzer according to their needsOr semi-automatic The instrument can also be set as a carbon analyzerCarbon analyzer

Instrument execution standard:

comply with national standardsGB/T15460-1995 "Determination of carbon and hydrogen in coal - Electricity gravimetric method"

Main technical parameters:

1Electrolytic voltage: Coating voltage:10V±0.5V

Working voltage:24V±0.5V

2Current: maximum electrolysis current: ≥800MA

Terminal current: 50mA ± 1MA

3Measurement range: Carbon:1%~100%

Hydrogen:0.5%~10%

4Sampling quantity:70~75mg

5Measurement repeatability: Carbon: ≤0.50% (Complies with GB476)

Hydrogen: ≤0.15% (Superior to GB476)

6Measurement time:10~15min

7Two stage furnace temperature:

combustion furnace850℃±10℃

reformer300℃±10℃

8. Power supply: Voltage 220V ± 22V Frequency 50HZ±0.5HZ

9. Host power: 280W Furnace power: 1500W

Main components: composed of host, computer, purification system, absorption system, cooling systemPt-P2O5Composed of electrolytic cells, etc.