Carbon hydrogen analyzer

NegotiableUpdate on 03/13
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The KY-300 analyzer embodies the advantages of various domestic hydrocarbon analyzers, and its precision and accuracy of analysis results meet the requirements of national standards GB/T 476 and GB/T 15460. It can be widely used in coal, power fuel, and electricity.
Product Details

KY-300Carbon hydrogen analyzerIt is a fully automatic system in China, which uses Coulomb method as the basic method and single-chip microcomputer technology to achieve intelligent and automated analysis of hydrocarbons. This instrument embodies various advantages in China, and the precision and accuracy of its analysis results comply with national standardsGB/T476andGB/T15460The requirements of this type of instrument can be widely applied to sectors such as coal, power fuel, electricity, economy and trade, scientific research, and petrochemicals. It also has the feature of one machine for multiple uses. Users can set the instrument as a hydrogen meter, carbon meter, or other instruments according to their needs.

Features:

vRealize multi-purpose use of one machine; Automatic data processing

vHigh measurement precision and accuracy

vWide measurement range; Automatically print analysis results

vThe reaction temperature can be freely set as needed

vSelf provided electrolytic cell cooling device

vFast analysis speed; Automatic temperature control and timing

vTemperature compensation; Over-temperature alarm

Instrument execution standard:

vGB/T476 Method for determination of carbon and hydrogen in coal

vGB/T15460 Determination of Carbon and Hydrogen in Coal Electricity--gravimetric method

Technical Specifications:

vMeasuring range Carbon:199% Hydrogen:0.199.9%

vRepetitive error: Carbon: ≤0.50% Hydrogen: ≤0.15%

vTemperature control range: Combustion furnace: room temperature~1000

Converter: room temperature~500

Converter: room temperature~500

vTemperature control accuracy: Temperature setting value ±10vAmbient temperature 040

vPower supply: AC voltage:220±22V 

vfrequency Rate: 50±0.5Hz

vload Qi: oxygen

vNo strong magnetic field interference

vmerit Rate: 3200W