Product Overview Implementation Standards: GB7600, GB/T11133CYW-106 Trace Moisture Analyzer adopts the classic theory - Karl Fischer microcoulometry method; According to the law of electrolysis, the number of water molecules reacted is directly proportional to the number of charges. The instrument detects the number of charges involved in the reaction (Coulomb) and automatically converts it into the corresponding number of water molecules. Therefore, this method has extremely low testing accuracy and cost, and has irreplaceable advantages over other testing methods; Reliable determination of trace moisture in liquid, gas, and solid samples
Product Overview
Execution standards: GB7600, GB/T11133
The CYW-106 micro moisture analyzer adopts the classical theory of Karl Fischer microcoulometry; According to the law of electrolysis, the number of water molecules reacted is directly proportional to the number of charges. The instrument detects the number of charges involved in the reaction (Coulomb) and automatically converts it into the corresponding number of water molecules. Therefore, this method has extremely low testing accuracy and cost, and has irreplaceable advantages over other testing methods; Can reliably determine trace amounts of moisture in liquid, gas, and solid samples. The instrument displays the measurement electrode signal in the form of a bar chart, intuitively indicating the water content of the electrolyte and depicting the real-time variation curve of electrolysis speed over time. It has the characteristics of high sensitivity, high precision, high reproducibility, low power consumption and energy-saving design. It can be equipped with a built-in battery for portable measurement and is widely used in fields such as petroleum, chemical, power, pharmaceuticals, inspection, scientific research, and environmental protection.
The types of detectable substances include:
- Gasoline, hydraulic oil, insulation oil, transformer oil, turbine oil, fire-resistant oil.
- Pentane, hexane, dimethylbutane, octane, dodecane, eicosane, octadecane, cyclododecane, decylcyclohexane, methylbutadiene, benzene, toluene, xylene, ethyltoluene, diethylene, tetracene, petroleum ether, cyclohexylamine, methylcyclohexylamine, cyclohexane, ethylcyclohexylamine, dicyclopentadiene, dimethylnaphthalene, triethylene, benzene, dihydrophenanthrene, fluorene, methylene phenanthrene, isopropylbenzene, etc.
- Phenols, cresols, fluorides, chlorophenols, dichloros, nitrophenols, etc.
- Ether derivatives include diethylene glycol monomethyl ether, diethylene glycol dimethyl ether, polyethylene glycol diethyl ether, benzyl ether, fluorophenyl ether, iodobenzene ether, didecene ether, and diheptene ether.
- All alcohols, all halogenated hydrocarbons, all lipids, etc.
Product Features
- 320 × 240 dot matrix graphic LCD display screen, touch screen operation
- Real time depiction of the variation curve of electrolysis speed over time
- Display the measurement electrode signal in the form of a bar chart to visually indicate the water content of the electrolyte
- Use blank current compensation and balance point drift compensation to correct measurement results
- Switching constant current electrolysis technology reduces overall power consumption of the machine
- Historical records with time stamps, up to 255 stored
- Equipped with electrode open circuit and short circuit self-test alarm function
- Built in high-speed thermal micro printer, with beautiful and fast printing, and offline printing function
- Built in battery (optional), can be used continuously for more than 6 hours after being fully charged
- Equipped with a standard RC232 interface, it can be connected to a computer for easy processing of test data
- Equipped with screen saver function to extend the lifespan of LCD
Product Parameters
| measurement range | 0ug~100mg |
| Accuracy | The error of testing water volume between 3ug and 1000ug is less than ± 2ug |
| Test water volume greater than 1000ug with an error of less than ± 0.2% |
| resolution | 0.1ug |
| Electrolytic current | 0~400mA |
| standby power consumption | 6W power consumption: 35W |
| power supply voltage | AC220V±20% 50HZ±10% |
| Applicable ambient temperature | 5℃~40℃ |
| Applicable environmental humidity | ≤85% RH |
| Overall dimensions | 350×260×180(mm) |
| weight | About 8kg |