Product Overview: The Karl Fischer titration method is used to determine trace amounts of moisture in different substances

Product Details



Product Overview

Using Karl Fischer titration method to determine trace water content of different substances. The micro moisture analyzer has successfully applied this method, using an automatic control circuit, a 32-bit embedded microprocessor as the main control core, and embedded in a mini operating system. This makes the instrument more reliable and convenient to use. It has the characteristics of fast analysis speed, simple operation, high accuracy, and strong automation. Widely used in sectors such as petroleum, chemical, power, railway, pesticide, medicine, and environmental protection.




Product Features

  • Adopting a 32-bit embedded microprocessor as the main control core, embedded with a mini operating system.
  • Constant voltage detection, high accuracy, fast measurement speed, stable and reliable.
  • Color touch screen, full numeric keyboard, easier operation, convenient and fast data calculation.
  • Contains 4 calculation formulas to meet customer needs.
  • Micro thermal printers come with time records for easier searching.
  • A brand new appearance designed according to ergonomics.
  • Innovative and optimized operating software, providing a unique touch experience.



Technical Specifications

  • Titration method: Electric titration (Coulomb analysis)
  • Display: Color LCD touch screen
  • Electrolytic current control: 0-400mA automatic control
  • Measurement range: 0ug~200mg
  • Sensitive valve: 0.1 µ g H2O
  • Accuracy: 10 µ g to 1000 µ g ± 3 µ g; no more than 0.3% for 1mg and above
  • Printer: Micro thermal printer
  • Power supply: 220V ± 10%, 50Hz
  • Power: < 40W
  • Operating environment temperature: 5-40 ℃
  • Operating environment humidity: ≤ 85%