1. Overview:
The CO2 incubator is a type of incubator that adds a CO2 inlet and a gas bottle to input CO2, meeting the environmental requirements of the culture. The Jinlan CO2 incubator series uses double-layer doors, LCD screens, glass doors for dehumidification, and stainless steel inner tanks, which have good heating, humidification, and insulation effects. It is mainly used for the cultivation of microorganisms, the cultivation of animal and plant cell tissues, etc. It is widely used in the production and research of cellular immunity, medical research, genetic genetics, biopharmaceuticals, environmental monitoring, agriculture and animal husbandry, and other biological engineering.
2、 Technical parameters: | model | | |
| SQ80L-IIQ | SQ160L-IIQ | SQ80L-IIW | SQ160L-IIW | CO80L-IIIQ | CO160L-IIIQ | CO80L-IIIW | CO160L-IIIW |
| Work area volume (L) | 80 | 160 | 80 | 160 | 80 | 160 | 80 | 160 |
| heating method | Gas heating type | Circulating water heat | Gas heating type | Circulating water heat |
| Temperature control range (℃) | RT+5~60 |
| Temperature resolution (℃) | 0.1 |
| Temperature fluctuation (℃) | ±0.1 |
| Unboxing recovery time (min) | (Using 37 ℃ as the detection standard) < 9min |
| CO2 control method | manual regulating valve | Imported infrared sensing probe, PID microcomputer |
| CO2 control concentration | 0-20% |
| CO2 recovery time | ≤5min |
| Humidification method | Natural evaporation of water pan |
| Whole machine power (W) | 450 | 720 | 680 | 950 | 450 | 720 | 680 | 950 |
| Partition board (block) | 2 |
Size of workspace (mm) (W*D*H) | 400*400*500 | 500*500*650 | 400*400*500 | 500*500*650 | 400*400*500 | 500*500*650 | 400*400*500 | 500*500*650 |
Overall dimensions (mm) (W*D*H) | 550*600*900 | 650*700*1050 | 550*600*900 | 650*700*1050 | 550*600*900 | 650*700*1050 | 550*600*900 | 650*700*1050 |
| power supply voltage | 220V 50Hz |
| Note: | |
3、 Product features:
1. The shell is made of high-quality steel, which undergoes quenching, acid washing, and leveling. The surface is electrostatically sprayed with a thick coating that is resistant to rust and corrosion. The overall thickness is finer, the surface is flat and smooth, beautiful, and has mechanical properties. The interior of the studio is made of mirror stainless steel, with rounded inner corners that are safe, touchless, and easy to clean. Partition argon arc welding, high strength, movable bracket, adjustable spacing up and down.
2. PID temperature control system, precise and easy to control temperature, with small fluctuations.
3. Double layered door structure, magnetic door seal, outer door with heating function, heating temperature can be set according to needs to effectively prevent condensation of the glass door inside the box. Reduce gas loss. The inner high-strength glass door allows for timely observation of the condition of the culture.
4. When the inner door is opened, the equipment automatically cuts off gas, the circuit stops running, and the gas supply valve closes, improving equipment safety while avoiding external pollution to the workspace.
5. The rear mounted ultraviolet sterilization lamp can be manually operated to regularly disinfect the box, effectively preventing contamination during cell culture and effectively removing bacteria and dirt.
6. High speed centrifugal fan, low noise, long continuous working time.
7. Manually adjust the valve to facilitate the control of the intake volume and concentration according to experimental requirements.
8. Gas heating method, fast heating rate, short recovery time, suitable for frequent opening of the box door.
9. Infrared sensor detection, unaffected by temperature and humidity, provides more accurate data.
10. Optional high-efficiency filters, USB and RS232 data interfaces can be used to connect to printers or computers.
4、 Attention:
1. When transporting the incubator, do not invert it and do not lift the door to avoid deformation.
2. Without injecting water before use, the device cannot be turned on, which may damage the heating element. Firstly, close the drainage valve and connect the water source to the inlet valve (provided by the water pipe user) to check if each connection interface is tight. Then turn on the power switch and water valve in sequence, observe whether the water level indicator light of the instrument controller is on, until there is a slight overflow of water on the side of the box, that is, cut off the water source.
3. Install the CO2 pressure reducing valve onto the CO2 cylinder, and there should be no air leakage at the interface. Do not open the cylinder valve temporarily. Connect the pressure relief output connector to the CO2 intake port on the back of the CO2 box with a rubber hose, and tighten it with a clamping ring to prevent air leakage.
4. The indoor environment should be well ventilated, shaded, and cool, with a distance of about 10 centimeters from the wall.
5. The power socket of the equipment should be reliably grounded to ensure safe use.
6. The operating room is equipped with a fan. Be careful not to insert fingers or foreign objects into the cover to avoid damaging the fan and causing safety accidents. Please cut off the power when swapping.
7. To maintain the appearance of the equipment, please do not use acidic or alkaline or corrosive substances to wipe the surface. The box can be cleaned regularly with a dry towel.
8. The rear of the equipment control box is equipped with a fuse. If the equipment is not powered on, the fuse tube should be checked for integrity first. When checking and replacing the fuse, please cut off the power supply. And replace the fuse with the same model.
9. When the low water level is insufficient, water should be replenished to the tank in a timely manner. Otherwise, the heating will stop. This water jacket heating type carbon dioxide incubator.
10. Stop using, please turn off the power switch. When not in use for a long time, please drain the water from the water jacket. This water jacket heating type carbon dioxide incubator.
11. The over temperature alarm indicator light is on and emits a sharp alarm sound. At this time, the power should be turned off for 30 minutes; If the temperature still exceeds the limit after turning on the power switch again, the power should be turned off and maintenance personnel should be reported.
12. When the pressure of the gas cylinder is lower than 0.2mpa, the cylinder should be replaced.
13. When the equipment is not in use for a period of time, before stopping work, the moisture in the working room must be removed, and the glass door must be opened for ventilation for 24 hours before closing.