E-mail
15046113@qq.com
-
Phone
13303065213,13315772373
-
Address
Wuxi City, Jiangsu Province
Wuxi Simjie Instrument Equipment Co., Ltd
15046113@qq.com
13303065213,13315772373
Wuxi City, Jiangsu Province
Asphalt needle penetration water bathSuitable for measuring the penetration of residual substances after distillation of petroleum asphalt, liquid petroleum asphalt, or evaporation of emulsified asphalt.with100gload5Insert the asphalt sample vertically within seconds to determine its hardness.The instrument consists of a circuit system, a load linkage, a standard needle, and a time control system.
Verification conditions for asphalt needle penetration ratio in water bath
3.1Equipment and instruments used for verification
3.1.1The division value is0.02 mmThe vernier caliper and division value are0.001mmThe outer diameter micrometer.
3.1.2The division value is0.01gThe electronic balance.
3.1.3The division value is0.1A stopwatch in seconds.
3.1.4The division value is0.1A thermometer with a temperature of ℃.
3.2The measuring instruments used for verification must pass the metrological verification and be within the valid period of metrological verification.
3.3The petroleum asphalt penetration tester should be used in15~30Verify at room temperature of ℃ and in a clean environment without corrosive gases.
Asphalt needle penetration water bathtechnical indicators
1. Needle insertion depth device: precision displacement measurement device.
2. Needle insertion time control device: adjustable within 0-99 seconds, with a time error of less than ± 0.1 seconds.
3. Needle insertion accuracy: 0.01mm.
4. The maximum needle insertion depth is 40mm.
5. Standard needle: The total weight of the needle and needle rod is 2.5 ± 0.05g, which meets the specifications of GB/T4509 and T0604.
6. Flat bottomed glass dish: It is a hard glass cylinder with a volume greater than 1.5L and a depth greater than 80mm. It is equipped with a sample dish holder inside to make it stable.
7. Sample container: made of copper, cylindrical with a flat bottom. The sample dish has a diameter of 55mm and a depth of 35mm (suitable for samples with a needle penetration of less than 200).
8. Lifting bracket: It can be freely adjusted to facilitate the alignment of the needle tip with the sample plane.