Smoke flowmeter operating condition parameter detector

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Smoke flow velocity meter working condition parameter detector This instrument uses the pitot tube method to measure the gas flow velocity in pipelines, and can detect various parameters such as smoke flow velocity, smoke flow rate, standard dry flow rate, dynamic pressure, static pressure, and smoke temperature in boilers, industrial furnaces, and exhaust ducts. The instrument uses an inlet high-precision sensor, and the sensor's 24-hour drift is less than 0.15Pa, making it particularly suitable for detecting low flow velocities

Product Details

GR-3020Type smoke flow velocity detector

Product Overview

Smoke flowmeter operating condition parameter detector The GR-3020 flue gas velocity detector (hereinafter referred to as the detector) is a portable monitoring device widely used in boilers, furnaces, and various exhaust ductsSmoke flow rate, smoke flow rate, standard dry flow rate, dynamic pressure, static pressure, and smoke temperatureMeasurement of parameters.

Scope of Application

Smoke flowmeter operating condition parameter detector This instrument uses the pitot tube method to measure gas flow velocity in pipelines, and can be used for various boilers, industrial furnaces, and exhaust ductsSmoke flow rate, smoke flow rate, standard dry flow rate, dynamic pressure, static pressure, and smoke temperatureThe instrument uses imported high-precision sensors for parameter detection, with a 24-hour self drift of less than 0.15Pa, especially suitable for detecting low flow rates.

adopt standards

JJG 518-1998Verification Regulations for Pitot Tubes

GB/T 16157 -1996Determination of particulate matter and sampling methods for gaseous pollutants in exhaust from fixed pollution sources

Main Features

1. Adopting imported high-precision micro differential pressure sensors, the pressure drift within 24 hours is less than 0.15Pa.;

2.The measurement accuracy of flow velocity is high, and the lower limit of determination can reach 0.3m/s;

3.Built in rechargeable lithium battery, capable of continuous operation for over 48 hours on a single charge;

4. Handheld measurement monitor, lightweight and portable, easy to operate;

5. Automatically calculate parameters such as average gas flow rate, average pressure, and flue gas flow rate.

6. Equipped with automatic zero point correction and software calibration functions to ensure measurement accuracy;

7.Equipped with flue gas distribution function, automatic sampling point count and measurement point distance;

8Large capacity data storage, capable of storing 800 sets of data files;

9Wide temperature LCD display with Chinese operating interface;

10.Large size, wide temperature, high brightness color display screen;

11.Equipped with power failure protection function, the sampling data will not be lost during power failure sampling;

12.Built in Bluetooth module, optional Bluetooth printer for data printing

working principle

Align the pitot tube with the airflow direction, with the negative end facing away from the airflow direction. When the flue airflow passes through the positive and negative nozzles of the pitot tube, a pressure difference will be generated. The microprocessor calculates the static pressure, flow rate, and air volume based on the collected dynamic pressure, total pressure, and smoke temperature signals, and then automatically calculates the * flow rate based on the input values of atmospheric pressure, humidity, pipeline cross-sectional area, and other parameters.

technical indicators

The main technical indicators of the flow velocity detector are detailed in Table 1.

Table 1 Main technical specifications of the detector

technical indicators

parameter range

resolution

accuracy

Smoke dynamic pressure

0~2000) Pa

0.01Pa

Not exceeding ± 2.0%

Smoke static pressure

(-35~ 35) kPa

0.01 kPa

Not exceeding ± 4.0%

flue gas temperature

0~600) ℃

1 ℃

Not exceeding ± 3 ℃

atmospheric pressure

(50~ 110) kPa

0.1 kPa

Not exceeding ± 4.0%

Smoke velocity

(0.3~45) m/s

0.1 m/s

Not exceeding ± 5.0%

Dimensions (length x width x height)

190mm×95mm×50mm

Continuous working hours

≥ 48 hours

power consumption

About 0.5W

Total weight

0.6kg

Product keywords: smoke velocity detector, high-precision smoke velocity meter, portable smoke velocity meter, flow velocity detector