Siemens CALOMAT 6 Thermal Conductivity Hydrogen Analyzer

NegotiableUpdate on 04/28
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Siemens CALOMAT 6 Thermal Conductivity Hydrogen Analyzer $r $nCALOMAT 6 Thermal Conductivity Gas Analyzer is mainly used for quantitative analysis of hydrogen or helium in binary gases or quasi binary mixed gases. If the thermal conductivity of certain gas components (such as Ar, CO2, CH4, NH3) differs significantly from that of other components, the CALOMAT 6 thermal conductivity gas analyzer can also be used to determine the concentration of these gas components in the sample.

Product Details

Siemens CALOMAT 6 Thermal Conductivity Hydrogen Analyzer

The CALOMAT 6 thermal conductivity gas analyzer is mainly used for quantitative analysis of hydrogen or helium in binary gases or quasi binary mixed gases.
If the thermal conductivity of certain gas components (such as Ar, CO2, CH4, NH3) differs significantly from that of other components, the CALOMAT 6 thermal conductivity gas analyzer can also be used to determine the concentration of these gas components in the sample.

application
• Pure gas monitoring (0-1% H2 in Ar)
• Protective gas monitoring (0-2% He in N2)
• Hydrogen monitoring (0-25% H2 in Ar)
• Synthetic gas detection (0-25% H2 in N2)
• Gas production
-0-2% He in N2
-0-10% Ar in O2
• Chemical applications:
-0~2% H2 in NH3
-50-70% H2 in N2
• Wood gasification (0-30% H2 in CO/CO2/CH4)
Blast furnace gas (0-5% H2 in CO/CO2/CH4/N2)
Acidic converter gas (containing 0-20% H2 in CO/CO2)
Equipment monitoring of hydrogen cooled generators:
-0-100% CO2/Ar in the air
- CO2/Ar 中0~100% H2
-80-100% H2 in the air
Explosion proof models are used for analyzing flammable and non flammable gases or vapors in hazardous areas

Order Data Order Number
CALOMAT 6 Hydrogen Analyzer - Single Channel 19 'Rack Unit Installed in Cabinet
Air connection: optional 6mm pipe or 1/4 'pipe

Measurement components (smallest measurement range)
H20 in N2... 1/100%
H2 0... 5/100% in N2 (blast furnace gas measurement) 1)
H2 0... 5/100% in N2 (converter gas measurement) 1)
H2 0... 5/100% in N2 (wood gasification) 1)
H20 in Ar... 1/100%
H20 in NH3... 1/100%
He in N2 0... 2/100%
He in Ar 0... 2/100%
He in H2 0... 10/80%
Ar in N2 0... 10/100%
Ar in O2 0... 10/100%
CO2 in N2 0... 20/100%
CH4 in Ar 0... 15/100%
NH3 in N2 0... 10/30%
H2 monitoring (turbine generator)
CO2 in the air 0... 100%
H20 in CO2... 100%
H2 in the air 80... 100%
Note: For other components, please contact Siemens
power supply
Optional 100 V... 120 V AC, 47... 63 Hz or 200 V... 240 V AC, 47... 63 Hz

Technical data
Overview (According to DIN EN 61207/IEC 1207, all data are obtained from H2 in binary mixture N2)
Range: 4 ranges, can be switched internally or externally, or automatic range can be used
Zui Large Range: 100% H2
Measurement during zero drift: Any point within the concentration range of 0-100% can be set as zero, with the smallest possible measurement range being 5% H2
Operation position: Vertical installation on the front panel
The standard complies with CE certification EN 61326/A1, EN61010/1

Design, chassis
Protection level IP 20, according to EN60529 standard
Weight approximately 10kg

CALOMAT 6 Thermal Conductivity Hydrogen AnalyzerElectrical Characteristics
EMC anti-interference (electromagnetic compatibility) (all signal cables must be shielded; strong electromagnetic interference can cause up to 4% of the maximum small range variation) complies with NAMUR NE21 (08/98) standard
Electrical safety complies with EN 61010-1, overvoltage level II
Power supply (see nameplate) AC 100-10%~120 V+10%, 47~63 Hz or AC 200-10%~240 V+10%, 47~63 Hz
Power approximately 20 VA
Insurance 100 .. 120 V: 1.0T/250; 200 . .. 240 V: 0.63T/250

Gas inlet conditions
Sample gas pressure is 80 110 kPa (absolute pressure)
Sample gas flow rate is 30 .. 90 l/h (0.5 ... 1.5 l/min)
Sample gas temperature 0 .. 50 °C
The temperature in the measurement room is approximately 60 ° C
Sample humidity<90% relative humidity

response time
Preheating time<30 minutes (reaching * accuracy after two hours)
Response time (T90)<5 seconds
Damping (electrical time constant) 0-100 seconds, programmable
Dead time (ventilation time for blowing the analyzer gas path at a flow rate of 1 l/min) is approximately 0.5 seconds