The pump suction xylene detector can detect the concentration of xylene in various environments. It adopts the PID photoionization principle sensor, which has high accuracy and can quickly measure the gas concentration on site. The range of 0-10ppm is generally used to measure the leakage concentration in normal air.

Product Details

pump-suction typeXylene detectorProduct Features

Adopting high-precision PID photoionization principle sensor

USB interface for high-speed data transfer, capable of downloading and printing data (optional function)

LCD dot matrix display technology can display gas types, gas units, gas concentration values, local time, etc

Supports both Chinese and English operation interfaces, with simple and convenient switching

Gas concentration units PPM, mg/m3 can be quickly switched for display

Built in miniature sampling pump with ten adjustable suction levels

Explosion proof certification, explosion-proof grade: Exia Ⅱ CT4

pump-suction typeXylene detectorProduct features

Pump suction sampling method, with adjustable suction force of the pump

One click factory reset function can eliminate the trouble of misoperation

Equipped with self calibration function and zero point calibration function, making gas monitoring more accurate and reliable

Equipped with temperature and pressure compensation, it can perfectly compensate for gas concentration under different temperature and pressure environments

Two level sound and light alarm, alarm points can be set by oneself

Large capacity lithium polymer rechargeable battery

The shell is made of high-strength special engineering plastic, which is anti slip, waterproof, dustproof, and explosion-proof

Exquisite and high-end aluminum alloy suitcase

Parameter Specifications

Gas Detected xylene
Detection principle PID photoionization principle
Sampling Method Pump suction, flow rate can reach 1L/min, the suction size of the pump has ten adjustable gears
Measure Range 0-1ppm、0-5ppm、0-10ppm、0-20ppm、0-50ppm、0-100ppm、0-200ppm、0-500ppm、0-1000ppm、0-3000ppm
Resolution Resolution 0.001ppm (0-1ppm), 0.01ppm (0-100ppm and below), 0.1ppm (0-200ppm, 0-500ppm), 1ppm (0-1000ppm and above)
精 度 Precision 2%FS~3%FS
Signal Output USB interface for high-speed data transfer, capable of downloading and printing data (optional)
Response Time ≤10S Recovery Time ≤10S
Repeatability ≤±1% Linearity error ≤±1%
Zero shift ≤± 1% (F.S/year) Display Technology LCD dot matrix display technology
Operating language Support both Chinese and English Gas unit Support switching display of multiple gas units
Working Temperature -20℃~50℃ Humidity at work 0-95%RH
Explosion proof rating ExiaⅡCT4 Alarm mode Sound, light, vibration

Environmental standards

Harmful substances in the air of workshops in China (TJ36-79)The maximum allowable concentration is 100mg/m3 (xylene)

Harmful substances in the atmosphere of residential areas in China (TJ36-79)Max allowable concentration0.30mg/m3 (primary value, xylene)

China (GB16297-1996) Comprehensive Emission Standard for Air Pollutants (Xylene)

maxAllowable emission concentration (mg/m3):

70 (Table 2); 90 (Table 1)

maxAllowable emission rate (kg/h):

Level 2 1.0 to 10 (Table 2); 1.2-12 (Table 1)

Level 3 1.5 to 15 (Table 2); 1.8-18 (Table 1)

③ Unorganized emission monitoring concentration limit:

1.2mg/m3 (Table 2); 1.5mg/m3 (Table 1)

Harmful substances in drinking water sources in China (to be promulgated)maxAllowable concentration 0.5mg/L (xylene)