Measuring principle
Using composite measurement techniques to separately measure the concentrations of ammonia nitrogen and nitrate in water samples. Ammonia nitrogen is measured using an ammonia sensitive electrode method, while nitrate is measured using a reagent free ultraviolet absorption method. Nitrate: Continuous spectral scanning of purple visible light (200-720nm), using absorbance at 220/275nm ultraviolet to calculate nitrate concentration. Easy to use, no need for reagents. Ammonia nitrogen: Add sodium hydroxide reagent to adjust the pH value of the sample to>12, and the ammonium ions in the water will be converted into ammonia gas. Use an ammonia sensing electrode to measure and convert the concentration of ammonia nitrogen. Anti chromatic interference, wide measurement range.
Application field
Drinking water
surface water
Domestic sewage
industrial wastewater
Simultaneous measurement of dual parameters, saving space and investment
Flow measurement, more stable
Chinese operation menu, easy to use
Fully automatic analysis, automatic calibration, and automatic cleaning
Optional outdoor floor mounted models
7-inch color touch screen Chinese menu, user-friendly human-machine interface
Android operating system
Ultra large data storage
Administrator password protection
USB software updates and data downloads
measurement method |
Nitrate: UV spectroscopic absorption method |
Ammonia nitrogen: Ammonia sensitive electrode method | |
Measurement range |
Nitrate: 0~5/10/20/200 mg/l |
Ammonia nitrogen: 0~10/100/1000 mg/l | |
measurement type |
Manual or automatic interval analysis |
repetitiveness |
Better than 5% |
detection limit |
0.15mg/l NO3-N |
0.05mg/l NH4-N | |
display |
7 "color touch screen |
Simultaneously display the measured values of ammonia nitrogen and nitrate nitrogen | |
Operation Menu |
Chinese, other options available |
data storage |
2G large capacity storage space |
Digital Communication |
RS 232/485 |
Analog output |
2-channel 4-20 mA output |
Protection level |
IP 65 |
power supply |
220VAC, 50/60Hz |
size |
600*500*300mm(H´W´D) |