High precision low turbidity measuring instrument, usage method of turbidity meter

NegotiableUpdate on 01/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

High precision low turbidity measuring instrument, turbidity meter usage method M-5010 High precision low turbidity measurement [Working Power Supply] 90 to 264 VAC, 50/60 Hz. 417 * 533 * 171 mm (W * H * D) Sedimentation tank effluent, filter effluent, factory effluent, etc

Product Details

Product Overview

The working principle of a low range turbidity online analyzer: When the measured water sample flows through a through type defoamer, the bubbles in the water sample are separated. When a parallel beam composed of a constant electric light source passes through the measured water sample, the suspended solids in the water sample emit scattered light, which is converted into a signal positively correlated with the suspended solids by a 90 ° direction photoelectric receiver. After processing by an electronic unit, the turbidity data of the water is obtained.

performance advantage

Suitable for monitoring turbidity in drinking water, pursuing low range and high accuracy

Principle of 90 ° scattered light measurement with stray light absorber

Glass bottle free module structure

Built in bubble elimination system ensures high precision and extreme measurement Excellent stability

Using LED light source does not require regular replacement of the light source

L Heating optical component design, suitable for low-temperature water quality

Low water consumption and fast response speed

Calibration package (350mL) calibration, simple and safe

Dry solid calibration material can be selected for verification

Compliant with US EPA standards

L has analog and digital communication outputs


Technical Specifications


range

0.0001 - 100 NTU

display

Color touch screen graphic display

accuracy

± 2% of the reading (0-10NTU)

calibration method

single-point calibration

detection limit

<0.0005 NTU

verification method

Standard liquid calibration or dry solid calibration material calibration

sensitivity

Better than 0.005 NTU

simulated current

1 channel 0/4-20mA, programmable

resolution

0.0001 NTU

relay output

3 alarm points SPDT, passive contacts

Repeatability/Accuracy

Better than 1% at 1 NTU


Digital Communication

Modbus RS485

Sample Temperature

0 to 50°C

Protection level

IP 65

Sample flow rate

30-500ml/minute, preferably 50-80ml/minute

meet the standard

US EPA

Environment/Operation

5 to 50 ° C, 5 to 95% (non condensing)

Hardware Installation

wall-mounted installation

Power requirements

90 to 264 VAC, 50/60 Hz.

size

417´533´171 mm (W´H´D)

pipeline connection

Inlet water - Ø 6 mm OD,

Water outlet - Ø 9 mm OD

weight

Less than 5 Kg