Overview

PH monitoring instrument product description $r $n -- It has two installation methods: online and submerged $r $n -- Automatic temperature compensation function $r $n -- Automatic configuration of PH or ORP $r $n -- Convenient connection method $r $n2760 preamplifier can be used in conjunction with rotary lock electrodes to connect Signet to any PH/ORP transmitter or monitor and other equipment.

Product Details

PH monitoring instrument GFpreamplifier:3-2760(new)

Installation type

old product

encoding

alternative product

encoding

description

immersive installation

3-2720

198864602

3-2760-1

159000939

preamplifier, NPT

3-2720-2

198864603

3-2760-2

159000940

preamplifier, ISO

Online installation

3-2720

198864602

3-2760-11

159001367

Online preamplifier, NPT

3-2720-2

198864603

3-2760-21

159001368

Online preamplifier, ISO


PH monitoring instrument2750The2760 DryLoctypePH/ORPSensor electronic components and preamplifiers

Measuring rangePH: 0~14 , ORP:±2000mV

temperature0~85℃

2750-1, 2750-2online

2760-11, 2760-21, 2760-31, 2760-41online

2750-3, 2750-4immersive

2760-1, 2760-2, 2760-3, 2760-4immersive

power supply: 2750 12-24VDC, 4-mAoutput

2760 6-12VDC

current output: 2750 PH:fixed4-20mAisolation=0-14PH

ORP:fixed4-20mA=-1000-2000mV

precision:±32uA