Characteristics of OM-CP-HITEMP140-FR Temperature Recorder
High temperature data logger with a 50 mm (2 ") fast response probe
Fast response time
Designed with food grade stainless steel, operating temperature range: -40 to 140 ° C (-40 to 284 ° F)
Optional ventilated or embedded insulation panels that can operate at temperatures up to 250 ° C (482 ° F)
The detection range is -200 to 260 ° C (-328 to 500 ° F)
Accuracy of ± 0.1 ° C (± 0.18 ° F)
Submersible (IP68)
Trigger setting
Programmable start and stop times
4 Hz reading rate
Ready to use type
Free NIST certification
OM-CP-HITEMP140-FR Temperature RecorderIt is a high-temperature data logger with ultra fast response time that can record temperature during rapidly changing thermal processes. This high-temperature data logger features a probe with a diameter of 50 L x 1.59 millimeters (2 x 0.0625 ") and is capable of recording frequencies up to 4 Hz, which is four times higher than other data loggers in its class. This enables temperature sensors to quickly adapt and accurately record temperature changes in constantly changing environments, with applications including oven contour analysis, steam * *, high-pressure * * validation and mapping, seafood processing, quick freezing, etc.
OM-CP-HITEMP140-FR Temperature RecorderDesigned with food grade stainless steel. The device can be placed in an environment with a height of 140 ° C (284 ° F), and the probe's measurement temperature range is -200 to 260 ° C (-328 to 500 ° F). OM-CP-HITEMP140-FR also offers an optional heat shield to extend the working range of the data logger to -200 to 250 ° C (-328 to 482 ° F). OM-CP-HITEMP140-FR-TSK (Heat Shielding Kit) comes with a ventilated or top flush housing, suitable for a variety of applications.
The "Trigger Settings" function of OM-CP-HITEMP140-FR allows users to configure high and low temperature thresholds, which will automatically start or stop recording data into memory when reached or exceeded. This data logger can store up to 32700 readings with date and time stamps, and has non-volatile solid-state memory that can retain data even when the battery is depleted.
OM-CP-MULTIMANT-Z is a multifunctional stand or bracket that can be used in conjunction with the OM-CP-HITEMP140 series data logger. It can be used to stabilize the recorder inside an autoclave, or it can be screwed onto a flat surface to form a fixed base. OM-CP-MULTIMANT-Z is made of 316 stainless steel and can withstand temperatures up to 150 ° C (302 ° F), making it ideal for high-pressure transformer processes.
OM-CP-HITEMP140-FR uses * new software. Data can be viewed in graphical or tabular form, and summaries and statistical views can be used for further analysis. This software has the ability to export to Excel ®, Data annotation, digital calibration, and other functions.
OM-CP-HITEMP140-FR Temperature Recorder Specifications
OM-CP-HITEMP140-FR
(Data logger without heat shield)
temperature
Temperature sensor: 100 Ω platinum RTD
Probe measurement range: -200 to 260 ° C (-328 to 500 ° F)
Temperature resolution: 0.01 ° C (0.02 ° F)
Calibration accuracy: ± 0.1 ° C (± 0.18 ° F) 20 to 140 ° C (68 to 284 ° F)
average
Data logger response time (hours: minutes: seconds: fractions of a second):
In the air:
T60-00:00:39:00 (60% of the step change can be achieved in 39 seconds)
T90-00:02:71:00 (2 minutes and 71 seconds, reaching 90% step change)
In the water:
T60-00:00:10:00 (60% step change can be achieved in 10 seconds)
T90-00:00:12:00 (90% of the step change can be achieved in 12 seconds)
Reading rate: 4 readings per second, 1 additional reading every 24 hours
Memory: 32767 readings
Startup mode:
• The software can be programmed to start immediately
• Delay * Start up 18 months in advance
Stop mode: manual or timed (specific date and time)
Trigger setting: Upper and lower limits can be set. Once the data reaches or exceeds the set limit, the device will record it in memory. Dual layer start and stop triggers can also be programmed. Users can read the number of readings after the device is triggered
Readings in trigger setting mode: 10922 readings
Real time recording: Can be used with a PC to monitor and record data in real-time
Password protection: Optional passwords can be programmed into the device to restrict access to configuration options. Read data without password required
Memory line break: Yes
Battery type: Comes with 3.6V high-temperature lithium battery; User replaceable
Battery life: Typical 2 years (1-minute read speed)
Calibration: Digital calibration through software
Calibration date: automatically recorded in the device
Data format: Date and timestamp in ° C, ° F, ° R, K
Time accuracy: ± 1 minute/month at 25 ° C (77 ° F)
Extension operation: ± 20 minutes/month at 140 ° C (284 ° F)
Computer interface: OM-CP-IFC400 USB docking station or OM-CP-IFC406 multiplexer interface is required; 125000 Porter
Software: Windows ® XP, SP3/Vista/7 and 8 (32-bit and 64 bit)
Working environment: -40 to 140 ° C (-40 to 284 ° F), 0 to 100% RH, 0.002 to 100 PSIA
Protection level: IP68
Size (body): 48 H x 24.6 mm D (1.89 x 0.970 ")
Size (probe): 44.5 L x 1.59 mm diameter (1.75 x 0.0625 "); 4.78 mm transition diameter (0.188")
Weight: 65 grams (2.3 ounces)
Material: 316 stainless steel, PEEK
OM-CP-HITEMP140-FR Temperature Recorder Ordering Guide
| Product Number | Product description |
| OM-CP-HITEMP140-FR | High temperature data logger with 50mm (2-inch) quick response probe |
| Additional purchase options | |
| OM-CP-IFC400 | Docking station with USB cable, software, and manual (for a single data logger) |
| OM-CP-IFC406 | A multiplexer data logger interface with USB cable, software, and manual (* can accommodate up to 6 data loggers) |
| OM-CP-BAT110 | Replace 3.6V high-temperature lithium battery |
| OM-CP-MULTIMOUNT-Z | Installation/bracket for OM-CP-HITEMP140 data logger |
| OM-CP-SVP-SYSTEM | IQ/OQ/PQ * * software validation workbook and software package compliant with FDA 21 CFR part 11 (unlimited users, license per computer) |
OM-CP-HITEMP140-FR Temperature RecorderComes with a 3.6V lithium battery. OM-CP-IFC400 software/cable packaging or OM-CP-IFC406 multiplexer (required for data logger operation)sold separately)Comes with an operation manual and USB interface cable.