MMC274Nexus ® Composed of two parts: the main unit (including electronic system) and a replaceable calorimeter module
MMC 274 Nexus® It consists of two parts: the main unit (including the electronic system) and a replaceable calorimeter module. This ensures maximum flexibility.
The instrument provides three replaceable calorimeter modules for different applications. Adiabatic acceleration calorimetry (ARC)®)The module is capable of conducting standard process safety tests in heating waiting detection and isothermal insulation modes. The Scanning module can perform temperature rise and constant temperature scanning tests similar to DSC, which can be used not only to study exothermic reactions but also to study endothermic reactions. The button battery measurement module is used for performance research on button battery charging and discharging and temperature scanning.
MMC 274 Nexus ™ Core functions:
- Multiple measurement modes are integrated, covering a wide range of application areas.
-Standard adiabatic tracking mode of heating waiting detection
-Isothermal adiabatic tracking mode
-Heating scanning calorimetry mode (constant power, linear heating)
-Isothermal scanning calorimetry mode
-Constant temperature difference mode, multi-stage mode, etc
- Based on the same desktop instrument, there are multiple calorimeter modules to choose from
-ARC module, used for security testing.
-Scanning calorimeter module, used for scanning tests similar to DSC.
-Button battery module, used for performance research of button batteries.
- Wide temperature range, up to 500 ° C
- Wide pressure range, up to 100 bar (10 Mpa)
- A variety of sample containers are made of different materials and offer different size specifications.
- Proteus® Software used to analyze starting point, peak temperature, area, etc. MMC data can be analyzed together with other thermal analysis data in the same interface.
MMC 274- Technical Parameters (Continuously Updated)
- Temperature range: 20 . 500°C
- Temperature resolution: 0.01 K
- Heating rate: 0 . 5 K/min
- Pressure range: 0 . 100 bar
- Pressure resolution: 0.01 bar
- Blast chamber capacity: 0.1 ml... 10 ml
- Adiabatic tracking rate: up to 50 K/min
- Tracking rate sensitivity: depends on temperature range, with a minimum of 0.002 K/min
|
|
Sample containers of different materials and sizes
|