Minolta CS-2000 spectrophotometer from Japan

NegotiableUpdate on 03/16
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Japanese Minolta CS-2000 Spectral Radiance Meter Minolta CS-2000 Spectral Radiance Meter Minolta CS-2000 Spectral Radiance Meter
Product Details

Specifications

Main specifications of CS-2000

model CS-2000
wavelength range 380 ~ 780 nm
wavelength resolution 0.9 nm/pixel
Display wavelength width 1.0 nm
Wavelength accuracy ± 0.3 nm (calibrated wavelengths: 435.8 nm, 546.1 nm, 643.8 nm, Hg Cd lamp)
Spectral wave width Below 5 nm (half wave width)
Measure angle (optional) 0.2° 0.1°
Measure brightness range (standard light source A) 0.003 to 5000cd/m2 0.075 to 125000cd/m2 0.3 to 500000cd/m2
Zui small measurement area 5 mm (1 mm when using a close-up lens) 1 mm (0.2 mm when using a close-up lens) 0.5 mm (0.1 mm when using a close-up lens)
Zui small measurement distance 350 mm (55 mm when using a close-up lens)
Zui small display brightness 0.00002 cd/m2
Accuracy: Brightness (standard light source A)*1 ±2%
Accuracy: Chromaticity (standard light source A)*1 x,y:±0.003(0.003 ~ 0.005 cd/m2
x,y:±0.002(0.005 ~ 0.05 cd/m2
x:±0.0015(0.05 cd/m2Above)
y:±0.001
x,y:±0.003(0.075 ~ 0.125 cd/m2
x,y:±0.002(0.125 ~ 1.25 cd/m2
x:±0.0015(1.25 cd/m2Above)
y:±0.001
x,y:±0.003(0.3 ~ 0.5 cd/m2
x,y:±0.002(0.5 ~ 5 cd/m2
x:±0.0015(5 cd/m2Above)
y:±0.001
Repeatability: Brightness (2 σ)*2 0.4 %(0.003 ~ 0.05 cd/m2
0.3 %(0.05 ~ 0.1 cd/m2
0.15 %(0.1 ~ 5,000 cd/m2
0.4 %(0.075 ~ 1.25 cd/m2
0.3 %(1.25 ~ 2.5 cd/m2
0.15 %(2.5 ~ 125,000 cd/m2
0.4 %(0.3 ~ 5cd/m2
0.3 %(5 ~ 10cd/m2
0.15 %(10 ~ 500,000cd/m2
Repeatability: chromaticity (2 σ)*2 0.002(0.003 ~ 0.005 cd/m2
0.001(0.005 ~ 0.1 cd/m2
0.0006(0.1 ~ 0.2 cd/m2
0.0004(0.2 ~ 5,000 cd/m2
0.002(0.075 ~ 0.125 cd/m2
0.001(0.125 ~ 2.5 cd/m2
0.0006(2.5 ~ 5 cd/m2
0.0004(5 ~ 125,000 cd/m2
0.002(0.3 ~ 0.5 cd/m2
0.001(0.5 ~ 10 cd/m2
0.0006(10 ~ 20 cd/m2
0.0004(20 ~ 500,000 cd/m2
Polarization error 1 °: less than 2% (400-780 nm); 0.1 ° and 0.2 °: below 3% (400-780 nm)
integration time Fast: 0.005-16 seconds; Normal: 0.005-120 seconds
measure time Zui 1 second less (manual mode) to Zui 243 seconds more (normal mode)
Color Space Mode Lvx y,Lvu' v',LvTΔuv,XYZ, Spectral curve, characteristic wavelength, excitation purity
interface USB 1.1
Operating temperature/
humidity range
5-35 ℃, relative humidity below 80%, no condensation
Storage temperature
/Humidity range
0-35 ℃, relative humidity below 80%, no condensation
power supply Power adapter (100-240~, 50/60 Hz)
power About 20 W
size 158 (width) × 262 (height) × 392 (length) mm
weight 6.2 kg
*1
Under the conditions of a temperature of 23 ± 2 ℃ and a relative humidity below 65%, measure the average value of 10 times in normal mode
*2
Measure 10 times in normal mode at a temperature of 23 ± 2 ℃ and a relative humidity below 65%

system diagram

系统配置

Data processing software

CS-s10w Professional Edition (Standard Accessories)

Using this software, CS-2000 can be measured through computer control and display data in various chart or list formats. Measurement data can also be copied, pasted, or exported to other software worksheets. CS-s10w provides a wide range of data management, analysis, and evaluation methods to help you conduct research and development or quality control work more easily and efficiently.


  • The above specifications are subject to change without prior notice.