Industrial PH/ORP meter PHG801PH online analyzer
This instrument only has one electrode input terminal, and PH should be installed when selecting PH mode Electrodes, OPR electrodes should be installed in ORP mode.
For measurement purposes, the shelf life of glass electrodes is one year. After one year from the factory, no Whether the pipe is used or not will affect its performance, and it should be replaced in a timely manner.
The pH electrode used or deactivated must be used before use Soak in a 3mol/L solution for 24 hours.
Please preheat the instrument for at least half an hour before performing calibration work.
Industrial PH/ORP meter PHG801PH online analyzer
■ Adopting a 2.4-inch 12864 dot matrix screen
■ Adopting isolated transmission output, with less interference
■ Adopting isolated RS485 communication
■ Can perform PH/ORP measurement, temperature measurement, upper and lower limit control, and variable Output transmission, RS485 communication
■ Configurable temperature manual and automatic compensation functions
■ High and low alarm functions, as well as hysteresis, can be set
■ Can set buzzer and LCD backlight switch functions
■ Add * password function
■ Industrial controlled watchdog ensures instruments do not crash