Welcome Customer !

Membership

Help

Suzhou Frog Vision Sensing Technology Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products
Product Categories

Suzhou Frog Vision Sensing Technology Co., Ltd

  • E-mail

    sales@broadsensor.com

  • Phone

    18610516470

  • Address

    Yida Technology Park, No.11 Jinpu Road, Suzhou Industrial Park

Contact Now

Fiber optic turbidity sensor

NegotiableUpdate on 01/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

PN:610211, 610211-1610211-2, frog vision turbidity sensor, based on the principle of 90 ° scattered light, using infrared LED light source and fiber optic as the main optical path, in compliance with ISO7027 international standard. The optical path and software processing only need to be 2 centimeters away from obstacles to remain unaffected. Can be used directly under outdoor sunlight.

Product Details

Advantages and Characteristics:

Using 90 ° light scattering method;
Only a small distance from the obstacle is needed to avoid its impact.
Using optical fiber as the main body of the optical path, the polishing and optical path processing, as well as software algorithms, have small drift and strong resistance to external light interference.
Easy to calibrate, only 50mL of standard solution is needed for on-site calibration.
Internal automatic temperature compensation.
RS485 digital output, supporting Modbus protocol.
Multiple range options are available to meet different needs.

name parameter
measurement method 90 ° scattering, ISO7027
measurement range 0.3~100NTU, 0.3~1000NTU, 1.0~4000NTU
resolution 0.01NTU, 0.1NTU, 1.0NTU
precision +/-3% FS
Operating temperature range 0~45℃
Storage temperature range -10~50℃
detection limit 0.3NTU
warranty 1 year
Using water depth IP68, 10m Max
power supply 5-12V 10mA (normal)
output RS485, Modbus RTU protocol
material
316, POM and PMMA optical fibers
size
187mm in length and 22mm in outer diameter
flow velocity < 3 m/s
response time 30 second T90
Service life* Sensor 2 years or more
Maintenance and calibration frequency* Sensor for 6 months
Note: * The lifespan, maintenance, and calibration frequency are for reference only and depend on the actual usage environment.

User Manual for Fiber Optic Turbidity Sensor

Click to download