An instrument used for automatic testing of MPO optical fibers.
MultiFiber ™ Pro Optical Power Meter and Fiber Testing KitThe only tester that does not use sector jumpers to test MPO optical cables, which makes polarity verification no longer complicated and makes MPO card boxes easier to conduct on-site testing. Whether using 10 pre terminated fiber optic cables or planning to use the new generation 40/100 Gbps performance, data centers are making MPO connectors the standard solution. A typical data center fiber optic installation means time-consuming, manual, and imprecise MPO verification. MultiFiber Pro is 90% faster than single fiber testing methods because it only measures power loss and polarity verification for 12 fibers in a single connector - reducing testing time from weeks to days.
The "All Scan" function in the MultiFiber Pro power meter automatically scans and tests all fibers in 12 MPO connectors - all loss and power measurements can be completed in just 6 seconds. This function can automatically test MPO terminal fibers when using sector jumpers, avoiding the process and time consumption of moving fiber testers between fibers.
The purpose of any polarity scheme is to provide an uninterrupted connection between the link transmitter and the link receiver. For array connectors, TIA-568-C.0 defines three methods to achieve this goal: methods A, B, and C. Deployment errors are common because these methods require combining jumper wires of different polarities for use. To ensure correct polarity, MultiFiber Pro allows users to test individual jumpers, links, and channels.
MPO connectors are installed on both the optical power meter and the light source, allowing testing of MPO fibers without the need for expensive and complex sector jumpers.
One of the main issues in data centers is whether it is possible to have a detailed understanding of the condition of a single fiber during testing and troubleshooting. MultiFiber Pro can diagnose faults in individual fibers in MPO optical cables and provide test results for each fiber. This feature can improve the flexibility of MPO connector validation and troubleshooting through more accurate single fiber data and reports.
Ports without protective covers pose a risk of contamination to wiring, equipment, and testing instruments, thereby affecting performance. Using connectors with shielded protective covers on the tester can prevent dust. By closing the MPO connector port, MultiFiber Pro provides you with higher testing reliability and peace of mind.
The MultiFiber Pro power meter can store 3000 test results internally (equivalent to 250 MPO cables), which can be uploaded to LinkWare for operation ™ 7 Cable Test Management Software on PC. LinkWare 7 can manage test results, edit cable IDs, print professional reports, and even export data to spreadsheet format.
MultiFiber Pro comes with multiple convenient toolkit configurations to meet all your cleaning, inspection, and validation testing needs. The toolkit includes:
Fiber Inspection - FiberInspector ™ Pro。 The T-fiber microscope is the ultimate visual inspection tool. Its 3.5 "screen can provide larger images through a double probe. The small probe can be held in the palm of the hand and switched between magnification of 250x and 400x by rotating the dial. FiberInspector Pro and optional MPO connectors are supplementary accessories to MultiFiber Pro testing components, providing a comprehensive kit for cleaning, inspecting, and testing MPO connected fibers and tapes.
Fiber Cleaning - IBC ® One click cleaner, cleaning work is very important. Fluke Networks' one click cleaner can immediately clean the fiber end face inside the connector. Simply push the tool tip to perform the cleaning operation and use the wheel to return to its original position - it's that simple. The MPO one click cleaner is bundled with the MultiFiber Pro toolkit, ensuring you have better cleaning tools from the beginning.
| General Specifications | |
| Operating Temperature | -10 º C to+50 º C |
| Storage temperature | -20 º C to+50 º C<> |
| Operating Humidity | 95% (10 º C to+35 º C) without condensation 75% (35 º C to+45 º C) without condensation Uncontrolled<10 °=''> |
| operating height | 4000 meters |
| Storage height | 12000 meters |
| vibration | Random, 2 G, 5 Hz to 500 Hz |
| Optical power meter (specified for use at 23 º C (73 º F) unless otherwise specified) | |
| Detector type | InGaAs |
| wavelength calibration | 850 nm、1300 nm、1310 nm、1550 nm |
| measurement range | 0 dBm to -50 dBm |
| Linearity of power measurement | ±0.1 dB1 |
| Uncertainty in power measurement | ±0.35 dB |
| Repeatability of power measurement | < 0.10=''> |
| Display resolution, dB or dBm | 0.01 dB |
| Power display unit | dBm、mW、μW |
| User selectable loss limit | 0.05 dB to 50.0 dB, with increments of 0.05 dB up to 10.0 dB, and increments of 0.1 dB up to 50.0 dB |
| Automatic wavelength detection | yes |
| polarity detection | Detect A, B, C and Corning Plug&Play ™ Polarity of Universal Systems |
| 2 kHz detection | yes |
| Record storage | 3000 records, each with one fiber optic cable (250 12 core optical cable) |
| external interface | USB 2.0, full speed |
| Optical connector | 12 core fiber MPO interface, no pin plug. Compatible with 50 μ m and single-mode fibers. The connector has an automatically closing protective cover. |
| power requirement | 2 AA alkaline batteries |
| battery life2 | >30 hours (typical) |
| Auto Close | 10. 20, 30, or 60 minutes (can be disabled by the user) |
| Low battery warning | The low battery icon flashes |
| size | 5.7 inches x 3.2 inches x 1.5 inches (14.5 cm x 8.0 cm x 3.9 cm) |
| weight | 10.9 ounces (g) 309 |
| 850nm light source (specified for use at 23 º C (73 º F) unless otherwise specified) | |
| emitter type | LED:850 nm |
| wavelength | 850 nm :±30 nm |
| Spectral Width (FWHM) | 850 nm: ± 50 nm (typical value) |
| output power | 850 nm: Greater difference between channels of ≥ 24 dBm+/-1dB |
| Power output stability3 | ≤± 0.1 dB for more than 8 hours |
| Encircled Flux | When the optical connector of the light source is 50/125 μ m, it complies with ANSI/TIA-526-14B, ISO/IEC 14763-3, and IEC 61280-4-1. |
| Optical connector | 12 core fiber MPO interface, no pin plug. Compatible with 62.5 μ m and 50 μ m. |
| pattern | 2 kHz modulated, automatic wavelength |
| power requirement | 2 AA alkaline batteries |
| battery life4 | >30 hours (typical) |
| Auto Close | 10. 20, 30, or 60 minutes (can be disabled by the user) |
| Low battery warning | The low battery icon flashes |
| size | 5.6 inches x 3.2 inches x 1.6 inches (14.2 cm x 8.1 cm x 4.1 cm) |
| weight | 10.9 ounces (g) 309 |
| Model | Instructions |
| MFTK1200 | The MultiFiber Pro testing basic toolkit includes the MultiFiber Pro optical power meter, 850 light source, test jumper, MPO adapter, and portable case |
| MFTK1400 | The MultiFiber Pro testing and inspection toolkit includes MFTK1200, FT600 FiberInspector Pro, and NF370 MPO endcaps |
| Model | Instructions |
| MFPOWERMETER | MultiFiber Pro optical power meter |
| MFMULTIMODESOURCE | MultiFiber Pro Multi Mode 850 LED Light Source |
| NFC-IBC-1.25mm | One click cleaner suitable for 1.25 mm LC and MU connectors and jumper wires (5 quantities) |
| NFC-IBC-2.5mm | One click cleaner suitable for 2.5mm SC, ST, FC connectors and jumpers (5 quantities) |
| NFC - IBC - MPO | OneClick cleaners for MPO/MPO connector (5 qty) |
| NFC-Kit-Case-E | Enhanced Fiber Optic Cleaning Kit - Includes one one key cleaner for 1.25mm connectors, one key cleaner for 2.5mm connectors, and one for MPO/MPO ® A one click cleaner for cleaning, a solvent pen, a cleaning tube, and a soft bag |
| FT600 | FiberInspector Pro |
| NF370 | FT600 MPO FT600 fiber optic microscope probe adapter |
| TRC-MPO-PP-B | 1 m test cord with MPO/MPO, Pinned/ Pinned, Type B polarity |
| TRC-MPO-UP-B | 1 m test cord with MPO/MPO, Unpinned/ Pinned, Type B polarity |
| TRC-MPO-UU-B | 0.3 m test cord with MPO/MPO, Unpinned/ Unpinned, Type B polarity |
| BKC-MPO-ULC | 1 m breakout cord for MPO Unpinned LC connector |
| BKC-MPO-USC | 1 m breakout cord for MPO Unpinned SC connector |
| ADP-MPO-B | Type B polarity MPO adapter |
| Model | Instructions |
| GLD-MF | MFPOWERMETER/MFMULTIMODESOURCE 1-year support |
| GLD-MFTK | MFTK1200, MFTK1400 1-year support |