Overview
CL541-PLUS calibrator, automatic thermocouple calibrator, analog output and measurement of 4 TC thermocouple types J, K, T and E plus mV, easy to read backlit display screen, easy to change TC thermocouple type and ratio
Product Details

Characteristics of CL541-PLUS calibrator

Automatic thermocouple calibrator
Simulate output and measure 4 types of TC thermocouples J, K, T, and E plus mV
Easy to read backlit display screen
Easily change TC thermocouple type and ratio
Use automatic output step rapid calibration
Protect with sturdy and durable rubber boots
Directly calibrate temperature (Celsius or Fahrenheit)
Compatible with all process instruments

useCL541-PLUS calibratorYou can check and calibrate all thermocouple instruments and measure thermocouple sensors. Automatically indicate the connections on the display screen for easy connection. Enter the store, factory, or site carefree. It adopts rubber sleeve and sturdy thin switch protection. Even in dark areas of plants with backlit displays, it is easy to operate. Stop carrying millivolt power supplies and thermocouple meters, you can calibrate directly at temperature (° C and ° F).

CL541-PLUS calibratorCan be used in conjunction with your T/C, including J, K, T, and E types. The adjustable digital potentiometer "DIAL" can quickly set any value within 0.1 ° C and store any three temperatures for instant calling. Quick calibration through automatic output step by step. Select steps 2, 3, 5, 11, and 21 to automatically increase the output by 100%, 50%, 25%, 10%, or 5%. Select the step times that match the system, which are 5, 6, 7, 8, 10, 15, 20, 25, 30, and 60 seconds, respectively. Directly connected to the smart transmitter, PLC, Thermocouple inputs for controllers and multi-channel recorders, and verify their outputs or displays. Measure and eliminate any faults in the thermocouple sensor connection, and locate any broken or high resistance connections. Connect sensors,CL541-PLUS calibratorThermocouples that automatically measure C or F degrees.

CL541-PLUS calibrator Specifications
(Unless otherwise specified, the rated temperature for all specifications is 3 ° C, 70% RH)
Accuracy: ± (0.008% of reading+0.006 mV) Cold end
Compensation: ± 0.1 ° F (± 0.06 ° C)
Millivolt range: -13.000 to 80.000 mV Operating temperature
Range: -25 to 60 ° C (-10 to 140 ° F)
Relative humidity range: 10% ≤ RH ≤ 90% (0 to 35 ° C), no condensation; 10% ≤ RH ≤ 70% (35 to 60 ° C), no condensation
Size: 143 L x 76.2 W x 40.64 mm H (5.63 x 3.00 x 1.60 ")
Weight: 12.1 ounces (including boots and battery)
Power: 4 "AA" alkaline 1.5V (inclusive)
Battery life: 50 hours
Low battery level: low battery level indicator, nominal operating time is 1 hour
Preventing incorrect connections: Overvoltage protection up to 60 Vdc (rated for 30 seconds)
Display screen: High contrast graphic LCD display screen. LED backlight is used in low light areas.

Input impedance:>10 megaohms
Open the thermocouple
Threshold: 10000 Ω nominal value
Pulse:<10 microampere pulse, lasting for 300 milliseconds
Normal mode suppression: 50/60 Hz, 50 dB
Common mode rejection: 50/60 Hz, 120 dB source
Output impedance:<0.3 Ω
Source current:>20 mA (driving 80 mV to 10 Ω)
Noise: ≤ 4 microvolts p-p for frequencies of 10 Hz or lower

Equipped with a rubber sleeve, 1 red and 1 black lead, with a retractable shielded banana plug and clip, 4 "AA" alkaline batteries, NIST calibration certificate, portable bag, and operation manual.