5500 pole method dissolved oxygen electrode

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Brief description: The 5500 pole dissolved oxygen electrode produced by GLI is used in conjunction with GLI's D63, D53, D33, and PRO-D3 dissolved oxygen analyzers or transmitters. Its feature is the use of an integrated replaceable membrane head component, which simplifies the tedious and time-consuming operation of replacing a new membrane. Due to each component including a pre installed film, electrolyte, and electrode, the replacement of the component will be able to replenish the consumables required by the sensor. Update time: July 29, 2017 Visit times: 314 Product model: 5540D0A

Product Details

  • detailed introduction
brand HACH/Hash

5500 pole method dissolved oxygen electrodeInstrument Description
Great Lake(GLI)He is a new member of Hash Corporation in the United States1970Since the beginning of the year30 For many years, we have been committed to providing online water quality analysis and detection instruments for water supply, sewage, and industrial process control. and1992Year by yearISO9001authentication.
GLIproduction5500 Series dissolved oxygen sensors andGLIofD63TheD53TheD33andPRO-D3Used in conjunction with dissolved oxygen analyzer or transmitter.
Its feature is the use of an integrated replaceable membrane head component, which simplifies the tedious and time-consuming operation of replacing a new membrane. Due to each component including a pre installed film, electrolyte, and electrode, the replacement of the component will be able to replenish the consumables required by the sensor. The sensor uses the Clark measurement technique with three electrodes. The third electrode has better long-term polarization, which makes the sensor more accurate and stable. The self-cleaning option includes a compressed air cleaning system that can be programmed to automatically clean the sensor film.

5500 pole method dissolved oxygen electrodeApplication field
1Mainly used for dissolved oxygen detection in urban and industrial wastewater treatment plants, typical applications include:

· Aeration tank

· aerobic/Anaerobic digestion tank

· Factory water

2It is also applicable for the determination of dissolved oxygen in various water bodies such as natural water, sewage, and saltwater.
working principle
5500 The working principle of the series dissolved oxygen sensor is to insert two metal electrodes into a small chamber filled with electrolyte, which is sealed with a selective thin film. Water and soluble ions cannot pass through this membrane, but oxygen and a certain amount of other gases and hydrophilic substances can pass through this thin film. By immersing the sensor in water, dissolved oxygen measurement can be performed. Due to the effect of the primary battery or external voltage, a potential difference is generated between the electrodes. This potential difference causes metal ions to enter the solution at the anode, while oxygen passing through the membrane is reduced at the cathode. The current generated by this is directly proportional to the transfer rate of dissolved oxygen through the membrane and electrolyte, and therefore the current is proportional to the partial pressure of oxygen in the water sample at a given temperature.
GLIof5500 The series dissolved oxygen sensor uses polarographic Clark cell technology and consists of a three electrode system: a gold cathode, a silver anode, and a silver reference electrode. Using a constant voltage to polarize the silver reference electrode for voltage stabilization results in a more stable potential compared to traditional dual electrode systems, as it does not generate sufficient interferenceD.O.Measured current. The voltage stabilization design of the reference electrode ensures long-term polarization stability throughout its service life, enablingGLISensors have higher accuracy and stability.

spare parts/attachment
Membrane head replacement component:

5500A1020-001All components include a pre installed semi permeable membrane, electrodes, electrolyte, andOCircle shaped.
replaceORing spare parts:

5H0675: Used for allGLIMembrane head, this oneOThe ring plays a sealing role between the sensor body and the membrane head component.
Calibration kit spare parts:

276M121012 A disposable calibration package, only used forSaturation methodAir calibration. These packages provide a stable atmosphere around the membrane head to provide high-precision calibration.
Independent air cleaning system:
276C3M1220-115(Used for)115 VACPower supply)
276C3M1220-230(Used for)230 VACPower supply)

Cleaning head air cleaning component

276G1230: Cleaning hair and 25 Feet(7.6 m)Composed of long air delivery pipes and quick release pipe installation accessories, excluding compressors and casings.
Attention: The cleaning head component requires an additional compressed air source.


The system provides compressed air to clean the membrane of the sensor. The system includes a cleaning head and 25 Feet(7.6 m)Long air delivery pipes, quick release pipe installation accessories, and installation inNEMA 4XThe compressor inside the casing.

technical indicators
Immersion materialNoryl®and316 Stainless steel body,PVCInstall adapter,Viton®OCircle,Teflon®MembraneNoryl®andRyton®Components, as well as nylon cable clamps and overflow risers
electrode material Gold cathode, silver anode, and silver reference electrode
temperature range32~1220~50
temperature compensationNTC 30 KOresistance
Zui low flow rate per second 0.016 Feet(0.5 cm
The biggest pressure145 psi10 bar
measurement range0.0~40.00 ppm
sensitivity Range of measurement±0.05%
repetitiveness Range of measurement±0.5%
Response time(20 130 Reaching a step change in seconds90%
film thickness 2 Inches(50 Microns)
Sensor cable (included) 15 Feet(4.6 m)Long cable with terminalMSType quick release plug (made of metal or plastic material)
Measuring principle Three electrode polarographic Clark battery technology
Electrical certificationCSAandFMauthentication,GLIofD53Type can be used forClass I,Div.2Groups A~D)The dangerous area

Selection Guide:

Note:
1When each sensor includes a replaceable membrane head assembly (with a factory pre installed semi permeable membrane inside), a protective cap, and12 A disposable calibration kit for highly accurateSaturation methodAir calibration.
2To order sensors with non-standard length cables, please indicate the corresponding model sensor's ordering number followed by a suffix“0A”Then note the total number of feet of the cable. Example:5540D0A0A25That means25 A foot long cable.