Y09-310 LCD AC laser dust particle counter

NegotiableUpdate on 03/05
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Brief description: The Y09-310LCD AC laser dust particle counter is a dust particle counter jointly developed by our factory and Nanjing University of Science and Technology. The control method has been upgraded from old-fashioned button based control to touch screen control, with an 8-inch color LCD touch screen (display size 162mm wide * 121mm high, display ratio 4:3), and a large visual interface making operation more convenient and intuitive.
Product Details
brand Other brands Price range 10000 to 30000
Instrument type laboratory Flow range large flow
Origin category domestic

1Y09-310 LCD AC laser dust particle counteroverview

This product is a dust particle counter jointly developed by our factory and Nanjing University of Science and Technology. The control method has been upgraded from old-fashioned button based control to touch screen control, with an 8-inch color LCD touch screen (display size 162mm wide * 121mm high, display ratio 4:3), and a large visual interface making operation more convenient and intuitive. The product adopts imported German Thomas air pump, which has lower noise and less electromagnetic interference; The exterior design is atmospheric and simple, with a small volume and light weight,Easy to use, with a fully stainless shell that better meets GMP requirements.

IIY09-310 LCD AC laser dust particle counterTechnical indicators:

model

Y09-310 LCDLaser dust particle counter

sample flow

28.3L/min ±5%

Particle size grading

0.3um,0.5um,1um,3um,5um,10um

laser light source

Long life laser diode with a lifespan of 10 years

Imported air pump

Continuous and stable operation, energy-saving and environmentally friendly

Verification standards

GB/T6167-2007,JJF1190-2008

Repetitive relative error

<±10%

Particle size distribution error

<±30%

Sampling concentration

35000 pieces/liter

cleanliness recovery characteristic

≤ 10 minutes (counting to zero for 3 consecutive times within 10 minutes, 95% confidence level)

sample time

User self setting (5-1800 seconds)

SAMPLE DELAY

User self set (1-100s) interval 1-1800s

UCL sampling frequency

2∽50

UCL sampling points (position)

2∽50

Data memory capacity

1 to 5000 sets of measurement data, searchable

count mode

Total number of

screen display

Color 8-inch touch screen

Printing method

Built-in Thermal Printer

power supply

Communication 220V/50Hz

Cleanliness level determination criteria

ISO14644-1, GMP Static, GMP Dynamic

Temperature and humidity

Optional

Data communication interface

USB

environment

Usage environment: 0-40 ℃, relative humidity 10-90%

environment

Standard accessory

Storage environment: -30 ℃ to 45 ℃, relative humidity 0 to 90%

environment

Standard accessory

External dimensions

Waiting for power sampling head, connecting hose (inner diameter 6mm) with self-cleaning filter, tripod, printing paper, portable protective box, power cord, USB flash drive, USB data cable, user manual, original factory calibration report, certificate of conformity

Length 315 X Width 260 X Height 230

material

SUS304 stainless steel

Weight (excluding packaging)

7Kg

IIIUsage conditions:

1. Clean room environment with no strong magnetic field interference around; Do not work in the atmosphere for a long time to prevent sensor contamination

2. Power supply: AC220V50Hz。

4after-sale service

One year warranty and free software upgrade