Y09-3016 Handheld Laser Dust Particle CounterIt is a newly developed handheld laser dust particle counter by the company that uses a full half body laser sensor. It can be connected to a PC computer data acquisition system for remote control and can directly observe the testing situation of the instrument. The test data can be analyzed and processed by the computer and saved as an Excel file. All technical indicators meet the requirements of the JJG547-88 verification regulations issued by the General Administration of Metrology. The entire machine adopts American microcomputer control processing technology, semiconductor laser sensor technology, and imported air pump technology, which has the characteristics of multiple functions, high measurement accuracy, fast speed, easy portability, and simple operation. The instrument can simultaneously measure the number of dust particles of multiple sizes in one sampling, and this product has been widely used in clean room testing of electronic production enterprises; On site inspection and leak detection of filters; It can monitor biosafety, HVAC systems, computer rooms, beverage packaging environments, drug and medical device production environments, hospital clean operating rooms, automotive spraying environments, microelectronics, biochemical products, food hygiene, fine chemicals, precision machinery and other production and research departments. It is an instrument for HVAC and pharmaceutical enterprises and their supervision and management departments to implement GMP standards and electronic production enterprises.
Technical Specifications:
| light source | Semiconductor laser with a lifespan of over 35000 hours |
| sampling flow | 2.83L/min (0.1ft3) imported sampling pump |
| display mode | LCD display, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, battery level, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter) |
| Rechargeable battery | Lithium ion battery, 7.4 V, 2200mAh |
| power supply | AC power adapter, AC: 100V~245V, 50/60 Hz to DC: 7.4V, 1A; |
| working hours | 6 hours (remaining battery indicator) |
| Counting mode | Cumulative value, difference value, concentration value |
| test method | Single, repetitive, continuous, computational, remote |
| unit conversion | The unit can be converted into m3 |
| Particle size channel | 0.3 、0.5、1.0、3.0、5.0、10μm, Simultaneously count six particle sizes |
| sampling period | 1~10000 (s) Delay count: 0~99 (S) Self purification time: ≤ 10 (min) |
| working environment | Temperature: 10-40 ℃ (50-104 ℉) Relative humidity: 20-90% RH, no condensation Atmospheric pressure: 86-106KPa |
| Temperature and humidity: | Choose (1) Temperature: 0-50 ℃± 1 ℃ (2) Humidity: 0-100% RH ± 5% |
| UCL settings | Sampling points (A): Set from 2 to 9 points. Sampling frequency per point: (L) 2-9 times Set measurement position: 0-999 |
| UCL report | Compliant with ISO14644-1, GB50073-2001, GMP, FS-209E standards |
| data storage | It can store 1000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, data location port), and the data will not be lost after power failure. |
| communication interface | RS232, 9600 baud rate |
| Alarm Settings | The instrument is equipped with a level alarm function, which can alert the cleanroom of levels 100, 1000, 10000, 100000, 300000, and 1000000 when the battery voltage exceeds the standard |
| Zero Count | Meets the requirements of JJG-547-88 Dust Particle Counter Verification Regulations JIS B9921:1997, with less than one count every five minutes |
| Overlap error | At 5%, 2000000 particles per cubic foot |
| Print function | External printer (optional) |
| Overall dimensions | 230×130×45(mm) |
| weight | 0.6kg |
| calibration | Traceable to the National Institute of Standards and Technology (NIST), our company has passed the metrological standard building assessment, and traceable. The Institute of Metrology and Measurement Technology can also conduct calibration on its own or by third-party metrology institutions |
| Standard Configuration | Suitcase/charger, equal power sampling head, zero filter, RS 232 connection cable, computer connection communication software CD |
| Optional parts | External printer, professional triangular sampling rack, temperature and humidity sensor |