Overview

The Prima series ultrapure water machine adopts the industry's largest acrylic crystal full touch screen panel, with 3-channel water quality monitoring, flow statistics, 5-level filter alarm, water leakage alarm, female voice reading information, perpetual calendar clock and other functions. The system is equipped with a 12 liter pressure water tank and an integrated filter element, making it the cost-effective choice! There are currently two sub product series of Prima-Q/S, which can produce RO reverse osmosis water, DI deionized water/high-purity water, and UP ultrapure water, with output of 15 liters and 30 liters per hour

Product Details

Product Introduction
The Prima series uses urban tap water as its water source and can easily and quickly manufacture RO reverse osmosis water, DI deionized water, and UP ultrapure water. It adopts a large-sized acrylic crystal full touch screen panel in the industry, with functions such as 3-channel water quality monitoring, flow statistics, 5-level filter alarm, water leakage alarm, female voice reading information, perpetual calendar clock, etc. It is equipped with a 12 liter pressure water tank.
There are currently two sub product series of Prima-Q/S, with a system water production capacity of 15 and 30 liters per hour, and an ultra pure water resistance rate of 18.2 M Ω. cm, which meets the Class I water quality standards established by GB/T 6682-2008, GB/T33087-2016, ASTM, CAP, CLSI, EP, and USP.
Performance Characteristics
Oversized (280 × 118mm) acrylic crystal full touch screen panel
Fully automatic microcomputer control system, multi-level menu operation
Online 3-channel water quality monitoring, real-time monitoring of source water, RO reverse osmosis water, DI deionized water/UP ultrapure water quality
Capable of traffic statistics and consumables management based on time and traffic
Equipped with a 5-level filter alarm prompt function, the filter life can be set, and automatic reminders for replacement upon expiration can be provided to avoid water quality deterioration
Standard female voice registration and alarm prompts, as well as sound and light alarm functions
No water alarm, full water alarm, leakage alarm, source water, RO reverse osmosis water, DI deionized water/UP ultrapure water water quality (parameters can be set freely) exceeding the standard alarm
The system comes with a perpetual calendar clock, which facilitates the setting of consumables and services
The fully automatic RO membrane anti scaling flushing program can set the flushing time and interval time to extend the service life of the RO membrane
The system has one built-in 12 liter pressure water tank, which saves laboratory space and makes installation and maintenance more convenient. An external large capacity water storage tank can be added to meet different water volume needs
Integrated plastic chassis, ergonomic design, water and electricity separation structure
Preprocessing RO、 The ultra purification component adopts a modular independent structure, making system maintenance and filter replacement more convenient and in compliance with GLP standards
Pure water pipelines and fittings have obtained NSF certification
Ultra long life composite KDF pre filter column, which can achieve 1 year without replacement and reduce operating costs
Optimized RO membrane module design, using Dow DOW * RO membrane from the United States, achieves the combination of long lifespan of RO membrane and high-quality water quality
A brand new integrated 4-column ultra purification column module that can be independently disassembled, using Dow * nuclear grade resin from the United States to ensure the quality of pure water at all times
Dual wavelength (185nm&254nm) UV lamp components (imported lamp tubes) effectively sterilize, reduce TOC, and enhance system applicability
MWCO5000DUF ultrafiltration module (*) effectively removes pyrogens (endotoxins) and can be used for precision cell culture and IVF
(0.45+0.1) μ m imported PES polyether sulfone composite membrane terminal sterilization filter, ensuring sterile water quality

Technical Specifications

name Basic type Heat source removal type Low organic matter type Comprehensive type
Product Model standard form Prima-S15 Primar-S15UF Prima-S15UV Prima-S15UVF
High traffic type Prima-S30 Prima-S30UF Prima-S30UV Prima-S30UVF
Water inlet requirements* Urban tap water: TDS<200 ppm, 5-45 ℃, 1.0-4.0 kgf/cm2When the TDS of the incoming water exceeds 200ppm, it is recommended to choose an external softener
System Process** PF+KDF+AC+RO+AC+DI+TF PF+KDF+AC+RO+AC+DI+UF+TF PF+KDF+AC+RO+UV+AC+DI+TF PF+KDF+AC+RO+UV+AC+DI+UF+TF
UP Ultra Pure Water Index:
resistivity 18.2MΩ .cm@25 ℃
heavy metal ions < 0.01 ppb
Total Organic Carbon (TOC)*** <10 ppb <3 ppb
bacteria <0.01 cfu/ml
Heat source (endotoxin) N/A <0.001 Eu/ml N/A <0.001 Eu/ml
Particulate matter (>0.2 μ m) <1/ml
Ribonucleases (RNases) N/A <1pg/ml N/A <1pg/ml
Deoxyribonucleases (DNases) N/A <5pg/ml N/A <5pg/ml
DI deionized water index
resistivity >5MΩ.cm
RO reverse osmosis water indicators:
Ion retention rate 96% -99% (when using a new RO membrane)
Organic interception rate >99%, when MW>200 Dalton
Particle and bacterial retention rate >99%
Water production rate (25 ℃)**** S15: 15 liters per hour; S30:30 liters/hour
Instantaneous water output 2.0 liters per minute (requires a pressure water tank), (adding UF models will reduce flow rate)
Water outlet (touch button) 3: RO reverse osmosis water, DI deionized water, UP ultrapure water
External dimensions/weight Length x Width x Height: 50 x 36 x 54cm/approximately 25Kg
power 220V、50Hz/72W
Standard Configuration Host (including 1 set of purification column)+15 liter pressure water tank+accessory package

*The quality of incoming water will affect the quality of pure water and the lifespan of the filter column

**PF: Pre filtration KDF: Copper zinc alloy AC: Activated carbon RO: Reverse osmosis DI: Ion exchange UV: UV lamp sterilization and digestion UF: Ultrafiltration TF: Terminal microfiltration

***Depends on the quality of the incoming water

****Inlet TDS=200ppm, 25? C. Measured at 50psi and 15% recovery ratio