TSI4000/4100 series laboratory and universal mass flow meters

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The 4000/4100 series quality flow meters produced by TSI are suitable for various gas flow measurement applications. Whether measuring airflow in laboratories or manufacturing workshops, TSI's universal mass flow meters can provide accurate measurement results and have multiple data output options. TSI mass flow meters are accurate to 2% of the reading and have a larger range. The fast response time of 4ms ensures that it can accurately measure fluctuating airflow, especially suitable for closed-loop control systems and integrated into flow measurement equipment. In addition, TSI mass flow meters have extremely low pressure drop, reducing the back pressure of the pipeline network and its impact on the system to a minimum

Product Details

Overview of Laboratory and General Quality Flow Meter Products:
Laboratory and General Quality Flow Meter 4000/4100 Series

The 4000/4100 series quality flow meters produced by TSI are suitable for various gas flow measurement applications. Whether measuring airflow in laboratories or manufacturing workshops, TSI's universal mass flow meters can provide accurate measurement results and have multiple data output options.

TSI mass flow meters are accurate to 2% of the reading and have a larger range. The fast response time of 4ms ensures that it can accurately measure fluctuating airflow, especially suitable for closed-loop control systems and integrated into flow measurement equipment.

In addition, TSI mass flow meters have extremely low pressure drop, minimizing the back pressure of the pipeline network and its impact on the system

Product Features RS232 interface: provides digital output and option configuration
4ms rapid response Set the zero point and conversion for analog output
Accurate to ± 2% of the reading Set the flow trigger value for starting/stopping measurement
High ultimate load ratio Set the refresh rate for LCD display
Low pressure drop Set the sampling speed for both analog and digital outputs
Analog signal output of traffic flow Select gas calibration
Universal digital outputs for flow rate, volume, pressure, temperature, etc Choose standard or volumetric flow measurement
NIST Calibration Certificate Set display unit for 4140/4143 type: L/min or cm3/min
Provide LabVIEW driver calculate volume
Technical parameters of 4140/4143/4040/4043/4045 models
4140/41403 4143/41433 4040 4043 4045
Inlet/outlet diameter 6.4 mm 9.525 mm 22 mm
ISO taper
12.7 mm 19.1 mm
Gas calibration air O2、N2
(41403 also includes N)2O)
air O2、N2
(41433 also includes N)2O)
air O2、N2Air and O2The mixture of gases
LCD display unit L/min,Std L/min,cm3/min,Std cm3/min L/min,Std L/min,
flow measurement scope 0.01∼20 Std L/min 0 ∼300 StdL/min 0 ∼200 StdL/min 0 ∼300 StdL/min
error Air and O2± 2% of the reading or 0.005 Std L/min, larger value N2O, N2: ± 3% of the reading or 0.010 Std L/min, larger value Air and O2± 2% of the reading or 0.05 Std L/min, with the larger value being air/O2Mixed gas, N2: ± 3% of reading or 0.1 Std L/min, whichever is greater
response time 4ms (63% of full scale) 4ms (63% of full scale)
size 127 × 49 × 32mm 182 × 63 × 53mm
Volume measurement* scope 0.01∼99.9 L
error ± 2% of the reading
pressure measurement scope
50-199 kPa (pressure)
error ±1kPa
response time <4ms (63% final value, step change)
temperature measurement
scope 0∼50℃
error ± 1 ℃ (flow rate greater than 1 Std L/min)
response time <75ms (63% final value, step change)
analog output
0-10V DC, flow only. Adjustable span (via RS232)
digital output RS232
DC power input 7.5 ± 1.5V, peak current 300mA