-Working principle:
Surface Scatter 7sc High Range Online Turbidity MeterUsing the principle of scattered light monitoring, the sample flows into the tilted turbidity meter body at a flow rate of about 1 to 2 liters per minute, and the liquid overflows from the horizontal plane at the top of the turbidity meter body, forming a flat measurement plane. A strong beam of light is directed at a certain angle towards the surface of the liquid, and suspended particles in the sample emit scattered light. A detector placed above the liquid surface detects the scattered light at a 90 degree angle.
-Application industry:
Surface Scatter 7sc High Range Online Turbidity MeterSuitable for the drinking water/sewage/industrial treatment water industry, measuring the turbidity of water with high suspended solids content such as raw water and sedimentation tank effluent.
-Instrument features:
This high range turbidity meter uses Surface Scatter technology to reduce interference and ensure reliability and accuracy.
Optical components do not come into contact with water and do not require frequent cleaning of the probe, reducing maintenance.
The range is wide from 0 to 9999NTU.
Using corrosion-resistant materials, high-temperature samples can be tested.
Equipped with a reference calibration board, it is more convenient to understand the current status of the instrument.
Can be connected to standard controllers sc200 or sc1000.
| technical indicators | |
|---|---|
| Measuring range | 0~9999NTU |
| Accuracy: | ± 5% of the reading between 0-2000NTU |
| Or ± 0.1NTU, whichever is larger; | |
| ± 10% of the reading between 2000 and 9999NTU | |
| Sample flow rate: | 1.0~2.0L/min |
| Power requirements: | 220 VAC, 50 Hz, 40 VA |
| Sample temperature: | High temperature 0-70 ℃, normal 0-50 ℃ |
| Analog, digital, alarm signals: | Please refer to the technical specifications of the SC200 or SC1000 standard controller |