Settlement balance

NegotiableUpdate on 03/29
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The TZC series settling particle size analyzer (settling balance) adopts the classic natural settling principle, which can conveniently and accurately determine the average particle size, median diameter, particle size distribution, standard error, specific surface area and other data of solid particles. There are two specifications, please click on the image on the left for detailed introduction.
Product Details

Produced by Shanghai Fangrui Instrument Co., LtdTZC series sedimentation particle size analyzerThere are two models of settlement balance, TZC-4 and TZC-5. TZC-5 is a comprehensive upgrade and improvement of the original TZC-4 model based on the advantages of existing similar instruments at home and abroad, which has brought the performance of the settling particle analyzer to a new height.

The main improvements include:
① Optimized the design of a high-performance electromagnetic force balance principle force sensor, increasing the accuracy from 1mg to 0.1mg, significantly improving the sensitivity, accuracy, and repeatability of the instrument;
② Adopting STN wide-angle backlit LCD display;
③ Multiple optimizations have been made to the data processing software, which can measure the settling performance and small particles of the sample for a long time. Smoother calculations and more accurate results;
④ Adopting the data structure of *, the original database program is reorganized to achieve more stable and reliable performance.

The TZC series settling particle size analyzer (settling balance) consists of high-precision electronic balance, computer, and particle size data processing software. When the measured particles in the settling liquid settle onto the balance plate, the LCD screen displays the mass value, and the mass signal is simultaneously transmitted to the computer. The particle size data processing software collects the mass signal in real time and displays it on the computer screen. After the settling is completed, the curve can be stored for easy retrieval, and then the particle size calculation can be performed. The calculation results can be printed out in the form of tables and graphs.

The TZC series sedimentation particle size analyzer is widely used to determine the particle size and distribution of dust in industries such as latex, pigments, grinding materials, synthetic resins, cosmetics, pharmaceuticals, glass, powder metallurgy, cement, plastics, refractory materials, coatings, petroleum, coal, etc. It is a commonly used instrument for scientific and technological personnel engaged in particle size research and testing work.

Main technical parameters:

model TZC-4 TZC-5
The largest load on the balance 100g 100g
Balance accuracy 1mg 0.1mg
Particle measurement range 1~600μm 1~600μm
Host function Zeroing, skinning, self calibration, alarm
display mode Wide angle backlit LCD display
data processing Collection, storage, and retrieval of settlement curves; Measurement of particle size, median diameter, average particle size, and specific surface area; average
Error calculation; Particle distribution analysis; Calculation results, chart printing

TZC series sedimentation particle size analyzer () standard configuration:
1 host (one self calibrating weight, two cables)
1 settling plate
2 ordinary settling cylinders
1 copy of user manual
1 certificate of conformity

TZC series sedimentation particle size analyzer options:
computer
printer

Testing principle of TZC settling particle size analyzer
According to Stokes' theorem, dust particles undergo particle classification during the settling process, and therefore the viscosity of the stationary settling liquid affects the settling
Particles play a role in frictional resistance, calculated according to the formula:
r = 9th/ [2 g (γk –γt)]? (H/t)
In the formula:
R=particle radius cm
η=viscosity of settling fluid, in grams per centimeter? Second
γ k=particle specific gravity g/cm3
γ t=specific gravity of settling fluid g/cm3
H=settlement height (distance from the settlement liquid level to the bottom of the weighing plate) in centimeters
T=settling time in seconds
G=acceleration due to gravity 980 cm/s2
After measuring the time t required for particles to settle to a certain height H, the settling velocity V and particle radius r can be calculated. The so-called settling analysis method applies this theory to obtain the distribution of particles.

Classic testing principles, clear and concise testing processes, stable and reliable testing data, fast and convenient data processing, simple and easy to understand operations, and low prices are particularly suitable for quality testing in industrial and mining enterprise laboratories, as well as for teachers and students in colleges and universities to carry out related scientific research.