Welcome Customer !

Membership

Help

Shandong Shengkang Electric Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products

SCKS402 type fully automatic open flash point tester

NegotiableUpdate on 01/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The open flash point tester measures the open flash point value of petroleum products when the test sample is used in an open environment (such as lubricating oil). Using touch screen instead of keyboard operation and adopting foreign * technology, the LCD large screen displays a human-machine dialogue interface in full Chinese, with full screen touch key prompts for input, which is convenient and fast. The open and fuzzy control integrated software has a modular structure and complies with national standards, the United States, the European Union, and other standards. It is an ideal substitute for imported instruments. Widely used in industries such as railways, aviation, electricity, petroleum, and scientific research departments.

Product Details

Instrument Features
Adopting a new high-speed digital signal processor, it operates reliably with high accuracy;
One host can simultaneously control multiple testing furnaces to conduct multiple sample tests, saving testing time;

The entire experimental process, including detection, lid opening, ignition, alarm, cooling, and printing, is automatically completed;

Platinum electric heating wire or gas ignition method (optional);

Automatic detection of atmospheric pressure and automatic correction of test results;

Adopting newly developed high-power high-frequency switching power supply heating technology, with high heating efficiency, and adopting adaptivePIDControl algorithm, automatically adjust the heating curve;

Temperature * automatic stop detection and alarm;

Thermal micro printer, making printing more beautiful and faster, with offline printing function;

Historical records with time stamps, stored as much as possible255a

A hundred year calendar clock with temperature compensation, accurate timekeeping, automatic recording of measured date and time, and can run in the event of power failure10Over years old;

adopt320x240Large screen graphicsLCDDisplay screen, Chinese character display interface, rich content;

Adopting full screen touch screen buttons, the operation is intuitive and convenient;

Multiple built-in execution standards are available for selection.

Can achieve full Chinese language proficiency/Full English interface display (optional)

Technical Specifications

● Temperature measurement:

Temperature measurement range: -59.9 ℃ -399.9 ℃

Temperature control range: Room temperature -399.9 ℃

Repeatability:150℃ 3℃

>150℃ 5℃

Resolution: 0.1 ℃

Accuracy: 0.5%

● Environmental temperature: 0-40 ℃

Relative humidity:<85% RH

Supply voltage: AC220V ± 20%

Maximum power consumption: 400W

Heating rate:

Compliant with national standards, US, EU and other standards

● Dimensions: 440X280X290 (mm)

● Instrument weight: 14kg