| brand | GREEN PRIMA/UK | price range | Negotiable |
|---|---|---|---|
| Classification | calcium ion | Origin category | Import |
| Instrument type | online type | Application field | Environmental protection, food, chemical, agricultural, pharmaceutical |

Online Hardness Analyzer _ UK GP New Product Launch
Accurate metering pumpTitration, maintenance free
Automatic calibration, automatic diagnosis, and monitoring
Low maintenance and reagent consumption
LCD display, with optional Chinese and English language menus
High cost-effectiveness and high measurement accuracy
Compact design, 400 × 300 × 180mm/5.5Kg
Continuous measurement or interval measurement (time adjustable)
All components in contact with liquids are made of corrosion-resistant materials
Multi channel programmable contact output
2 x Programmable Digital Input
Easy installation and simple operation
SD card data storage, fault recording, 4-20mA, RS485 output, etc
Water hardness tester, water hardness analyzer, online water hardness analyzer, imported water hardness tester, colorimetric water hardness analyzer
model |
PROCON8000 |
measurement method |
Titration colorimetric method |
language |
Chinese, English |
measure time |
About 3 minutes, depending on the hardness of the water |
accuratesex |
± 2% of the upper limit of the selected reagent |
repetitiveness |
± 1% of the upper limit of the selected reagent |
power supply |
230V or 240V ± 10%, 50-60Hz, power consumption: 25VA (during operation), 3.5VA (during standby) |
size |
300×400×180mm(W×H×D) |
display |
Graphic backlit LCD display |
optional unit |
°dH、°f、ppmCaCO3、mmol/l |
output |
3. 4 to 20mA signal. 4. RS485 5, CAN bus output |
Analysis cycle |
Time interval measurement (0-255min)/external signal/flow signal |
Rinse time |
Can be set (0-99s) |
water quality requirements |
pH:4-10; Iron:<3ppm; Copper:<0.2ppm; Aluminum:<0.1ppm; Manganese:<0.2ppm |
temperature |
Environment: 10 ℃ -50 ℃, water sample: 5 ℃ -40 ℃ |
pressure |
Ca 0.3-8 bar (max) (recommended 1-5 bar) |
Protection level |
IP65 |
Installation method |
wall-mounted |
Pharmaceutical water hardness tester _ Preferred UK GP
parameter |
reagent |
°dH |
°f |
ppmCaCO3 |
mmol/l |
Hardness reagent |
TH8001 |
0.05-0.50 |
0.09-0.89 |
0.89-8.93 |
0.01-0.09 |
Hardness reagent |
TH8002 |
0.25-2.50 |
0.45-4.48 |
4.47-44.7 |
0.04-0.45 |
Hardness reagent |
TH8003 |
1.0-10.0 |
1.79-17.9 |
17.9-179 |
0.18-1.79 |
Hardness reagent |
TH8004 |
2.5-25.0 |
4.48-44.8 |
44.8-448 |
0.45-4.48 |
Online hardness analyzer GREENPRIMA UKApplication fields:
Monitoring of water quality and hardness for boiler water, softener effluent, cooling circulating water, surface water, pharmaceutical injection water, etc;