Orion Dual StarTM desktop pH/ion concentration/millivolt meter

NegotiableUpdate on 05/05
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Orion Dual StarTM desktop pH/ion concentration/millivolt meter

Product Details

Orion Dual StarTM desktop pH/ion concentration/millivolt meter
Model: 58825-10,20
characteristic:
Place of Origin:

America

Instrument Introduction:

The Orion Dual StarTM pH/ISE dual channel instrument perfectly combines the Orion 720A+/920A+with the Orion Star Plus, making it easy to achieve

Simultaneously measure pH and ISE. Two independent input channels can simultaneously measure pH or ISE, and each has its own independent temperature compensation electrode interface that can be set

To display only one channel reading or simultaneously display dual channel readings, automatic blank can be selected in the low ion concentration range to ensure measurement stability, and it has

Single point or double point known addition or subtraction methods are used to measure ion concentration Directly control the stirring electrode (optional) and calibrate a single point in the curve

Can be edited separately to reduce errors, calibration curves and slopes can be displayed on the screen or printed, and can be connected to Orion automatic samplers (software)

Version 1.5 or above, relevant software can be downloaded for free to save measurement time and improve measurement accuracy under high sample flux

Orion EZ StartupTM can quickly and easily start the instrument, allowing users to set the instrument to the desired settings through a basic menu

Application mode. Built in endpoint judgment (SMART STABILITYTM) and endpoint display technology (SMART AVERAGINGTM) functions,

Optimized response time and measurement results. A GLP compliant memory can store: 1000 sets of measurement results, including time and date; all

Calibration data, as well as pre stored 10 sets of measurement methods for each channel (a total of 20), all calibration curves and data can be obtained through USB or bidirectional RS-232 ports

Output to computer or printer

When measuring two different samples simultaneously or measuring two different parameters for the same sample simultaneously, the Orion Dual StarTM pH/ISE dual channel meter

It can achieve perfect measurement. For example, pH and ammonia nitrogen concentration in wastewater can be easily and quickly measured in the laboratory, and pH and sodium ion concentration can be measured simultaneously in food inspection.

The instrument includes the following hardware interfaces:

Two BNC interfaces

Two reference electrode interfaces

Two automatic temperature compensation electrode interfaces

A mini USB interface

A 3.5mm stirring electrode interface

A 2.5mm RS232 interface

A 0.5mm power interface

A grounding wire interface

to configure:
Standard model: Host, quick boot card, can be filled with Ross Ultra ® Composite glass electrode, electrode holder, stainless steel temperature compensation electrode, universal transformer.
Complete set of models: host, quick guide card, can be filled with Ross Ultra ® Composite glass electrode, electrode holder, stainless steel temperature compensation electrode, 475ml of 4,7,10 calibration buffer solution
One bottle each of Ross ® One bottle of electrode storage solution, universal transformer.
Technical parameters:
pH
measuring range
-2.000 to 19.999 pH
resolution
0.1/0.01/0.001 pH
relative accuracy
±0.002 pH
Correction point
1-6 o'clock
Automatic buffer solution
US, NIST, DIN, and user-defined
ion concentration
measuring range
0 to 19900 M/ppm
resolution
1, 2, and 3 significant digits
relative accuracy
± 0.2 mV or 0.05% (whichever is larger)
unit
Ppm, ppb, M, %Or no unit
measuring method
Single known increase/decrease method; Double known addition/reduction method
Correction point
1-6 o'clock
Redox potential (ORP)
measuring range
±1999.9(mV / RmV / EH)
resolution
0.1mV
relative accuracy
± 0.2mV or 0.05% (whichever is larger)
temperature
measuring range
-15 to 105 ℃
resolution
-5 to 99.9 ℃ is 0.1 ℃; Above 99.9 ℃ is 1.0 ℃
relative accuracy
0.1℃
Instrument functions
LCD
Dual channel display
input
Two BNC interfaces, two reference electrode interfaces, and two MiniDIN interfaces (temperature probes)
Method storage
20 of them
Password protection
correct
Data Storage
1000 sets
Compliant with GLP standards
correct
data transmission
Bidirectional RS232 interface, optional USB cable
IP GRADE
IP54 dustproof and splash proof
Quality and Safety Certification
CE, CSA, UL, TUA, FCC Class A
operating environment
operate temperature
5 to 45 ℃
relative humidity
5 to 85% (non condensing)
power supply
4 x AA batteries, universal power adapter
External dimensions
9.4 cm (height) x 17.0 cm (width) x 22.4 cm (length)
Ordering Guide:
58825-10 Orion Dual StarTM desktop pH/ion concentration/millivolt meter, standard model
58825-20 Orion Dual StarTM desktop pH/ion concentration/millivolt meter, complete set of models
Market:
• Wastewater • Drinking Water • Life Science Research • Food and Beverage
Application:
• Emission testing • Chemical and reagent production • Process and quality testing • Product development • Finished product inspection