Overview

Measurement range: 0-5 mg/L Total cadmium/Total iron/Total nickel/Total lead/Total arsenic/Total zinc/Total copper/Fluoride/Total manganese/Total chromium Hexavalent chromium (range expandable) Accuracy: When the accuracy is ≥ 0.2 mg/L, it does not exceed; amp; amp; amp; amp; #177; 10%; When<0.2 mg/L, not exceeding amp; amp; amp; amp; #177; 0.02 mg/L。 Repeatability: not exceeding amp; amp; amp; amp; #177; 5% or not exceeding amp; amp; amp; amp; #177; 0.02mg/L measurement cycle: The small measurement cycle is 20 minutes, and according to the site, the color development time can be modified arbitrarily within 5-120min. Sampling cycle: Time interval (adjustable from 10-9999min) and whole point measurement mode. Calibration cycle: 1-99 days, adjustable at any time. Output: 4-20mA RS232. Environmental requirements: Temperature adjustable indoor, recommended temperature+5-28 ℃; Humidity ≤ 90% (non condensing) $n AC power supply: 220VAC amp; amp; amp; amp; #177; 10%, 50Hz $n abnormal alarm and power outage will not lose data $n touch screen display and command input $n abnormal reset and power outage after power on, the instrument will automatically discharge residual reactants from the instrument and automatically restore working status

Product Details

Online heavy metal analyzerCategory:

1XT-IVType online total cadmium water quality analyzer

Measurement method:TPPS4Spectrophotometric detection

Basic principle: After mixing the water sample with a regulator, the cadmium in the sample is exposed to alkaline environment and indicator(TPPS4)In the presence of the indicator, a colored complex is formed by reacting with it. The analyzer detects the change in color and converts it into a cadmium value for output. The amount of colored complex generated is equivalent to the amount of cadmium.

IIXT-IVType online total iron water quality analyzer

Measurement method: Phenanthroline spectrophotometric detection

Basic principle: Trivalent iron ions in water samples are reduced to ferrous ions by hydrogen amine hydrochloride. Under the action of buffer solution, ferrous ions react with ortho phenanthroline to form a colored complex. The analyzer detects the color change and converts it into total iron value for output. The amount of colored complex generated is equivalent to the total iron content.

IIIXT-IVType online total nickel water quality online analyzer

Measurement method: Diketoxime spectrophotometric detection

Basic principle: The water sample is oxidized by the oxidant ammonium persulfate, and in an alkaline environment and the presence of an indicator (dimethylglyoxime), it reacts with the indicator to form a colored complex. The analyzer detects the change in color and converts this change into a total nickel value output. The amount of colored complex generated is equivalent to the total nickel content.

4XT-IVType online total lead water quality analyzer

Measurement method:TPPS4Spectrophotometric detection

Basic principle: After mixing the water sample with a regulator, the lead in the sample is exposed to an alkaline environment and an indicator(TPPS4)In the presence of the indicator, a colored complex is formed by reacting with it. The analyzer detects the change in color and converts it into lead value output. The amount of colored complex generated is equivalent to the amount of lead.

5XT-IVType online total arsenic water quality analyzer

Measurement method: Arsenic molybdenum blue spectrophotometric detection

Basic principle: The water sample is first digested by acid at high temperature and then mixed with a regulator. The arsenic in the sample reacts with the indicator in an acidic environment to form a colored complex. The analyzer detects the color change and converts it into the total arsenic value for output. The amount of colored complex generated is equivalent to the total arsenic content

VIXT-IVType online total zinc water quality analyzer

Measurement method: Zinc reagent spectrophotometric detection

Basic principle: After acid digestion, the water sample is mixed with a regulator and then oxidized by an oxidant. In an alkaline environment and the presence of an indicator, it reacts with the indicator to form a colored complex. The analyzer detects the change in color and converts this change into the total zinc value output. The amount of colored complex generated is equivalent to the total amount of zinc.

7XT-IVType online total copper water quality analyzer

Measurement method: Sodium diethyldithiocarbamate spectrophotometric detection

Basic principle: This study uses the sodium diethyldithiocarbamate spectrophotometric method to determine total copper. At high temperatures, various copper compounds in the water sample are oxidized and digested by nitric acid to form divalent copper ions. Under ammonia conditions, copper ions react with sodium diethyldithiocarbamate to form yellow brown complexes. Within a certain concentration range, the absorbance of the system is proportional to the concentration of copper ions, and the total copper content in the water sample can be determined by spectrophotometry.

8XT-IVType online fluoride water quality analyzer

Measurement method: Fluorine reagent spectrophotometric detection

Basic principle: Fluoride ions in water samples react with fluoride reagents and lanthanum nitrate in acetate buffer solution to form colored complexes. The analyzer detects the change in color and converts it into fluoride ion values for output. The amount of colored complex generated is equivalent to the amount of fluoride ions.

9XT-IVType online total manganese water quality analyzer

Measurement method: High * method spectrophotometric detection

Basic principle: The water sample is dissolved at high temperature by an oxidant, and after cooling, it forms a colored complex with a color developing agent. The analyzer detects the change in color and converts this change into the total manganese value for output. The amount of colored complex generated is equivalent to the total manganese content.

10XT-IVType online total chromium water quality analyzer

Measurement method: Diphenylcarbazide spectrophotometric detection

Basic principle: After mixing water sample and strong oxidant for digestion, trivalent chromium is oxidized to hexavalent chromium. Hexavalent chromium reacts with the indicator in an acidic environment to form a colored complex. The analyzer detects the color change and converts it into the total chromium value for output. The amount of colored complex generated is equivalent to the total chromium content.

EleventhXT-IVOnline hexavalent chromium water quality analyzer

Measurement method: Diphenylcarbazide spectrophotometric detection

Basic principle: After mixing the water sample with a regulator, hexavalent chromium reacts with the indicator in an acidic environment to form a colored complex. The analyzer detects the color change and converts it into a hexavalent chromium value for output. The amount of colored complex generated is equivalent to the amount of hexavalent chromium.