Welcome Customer !

Membership

Help

Kemu Electronics (Shanghai) Trading Co., Ltd. - Kyoto Electronics China (KEM)
Custom manufacturer

Main Products:

mechb2b>Products

Kemu Electronics (Shanghai) Trading Co., Ltd. - Kyoto Electronics China (KEM)

  • E-mail

    0937982501@163.com

  • Phone

    13601731964

  • Address

    Room 1201, Huixin International Building, 333 Yishan Road, Xuhui District, Shanghai

Contact Now

Mixed Karl Fischer Moisture Analyzer MKH-710M

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The MKH-710M hybrid Karl Fischer moisture analyzer is a new product from Kyoto Electronics Industry (KEM) in Japan. It is also a Karl Fischer moisture analyzer that combines volumetric titration and coulometric titration. The Hybrid titration method, based on the design concept, can quickly and accurately measure the moisture content in samples ranging from 1ppm to 99%. Additionally, automatic titration calibration can be performed without the need for standard substances. The electrolytic calibration measurement system automatically starts the volumetric Karl Fischer reagent titration test with one click, eliminating human error and eliminating the need for pure water or standard substances to perform Karl Fischer reagent titration calibration. The Hybrid Karl Fischer Moisture Determination Method automatically performs parallel operations of volumetric titration and coulometric titration based on the moisture content in the measurement cell, providing high-speed and high-precision measurement results.

Product Details


Mixed Karl Fischer Moisture Analyzer MKH-710M

Mixed Karl Fischer Moisture AnalyzerMKH-710M Main Features:

1. technologyMixed titration method:Even if the moisture content in the sample is too high, rapid and high-precision testing can still be performed.
2. brand newElectrolytic calibration measurement system:Automatically perform titration calibration without human error, with high accuracy.
3. Integrated measuring pool:Capacity based Coulombic titration cupAnd separableEnsure sealing and improve convenience.
4. Intelligent burette:The titrant information is stored in the chip, and two sets of burettes can be installed simultaneously on the measuring unit.
5. Simultaneous measurement of four measuring units:The main control unit can connect four measuring units simultaneously and perform testing simultaneously.
6. Wireless Bluetooth transmission:The main control unit and measurement unit can be transmitted through a Bluetooth adapter for safer operation.
7. Electrode replaceable diaphragm occurs:detachableformulaicElectrode separator,Oil or contaminated samplesEasy to clean and maintain.
8. Three measurement modes:capacitytitrationLaw or Coulombpower levelBoth the method and the mixed method can be performed in the same measurement cell.
9. standard configurationUSBconnect the hole:useUDisk storage method and results as electronic records, and has the function of using permissions.

Mixed Karl Fischer Moisture AnalyzerMKH-710M Technical Specifications:

Coulomb electric quantity method
measurement range:水分1μg300mg H2Obromine index8μg300mg Br.
accuracy:RSDsmallYu0.3%(n=10)(measurement1mgWhen it comes to moisture).
resolution:0.1μg.
Control method:Constant current pulse time control.
Endpoint detection:Dual platinum detection electrode AC polarization method.
Terminal judgment:Drift stability determination, or setting measurement time.
容量滴定法
measurement range:水分100μg500mg H2O(According to Karl Fischer reagent titration degree).
moisture content:1ppm99% H2O.
Accuracy of burette: 10mL, Resolution1/20,000accuracy±0.015mLrepetitiveness±0.005mL.
Endpoint detection:Detection of polarization potential using dual platinum electrodes.
Terminal judgment:Determination of endpoint for duration of polarized potentialEnd time range199second.
Capacity Coulomb mixture method
Determination method:Karl Fischer titration methodandKarl Fischer Coulomb electric quantity method.
measurement range:1μg500mg H2O.
accuracy:RSDless than0.3%(n=10)(measurement10mgWhen it comes to moisture)(Karl Fischer reagent titration degree3mg/mL).
resolution:0.1μg.
Control method:Automatic switching of speed for continuous titration(容量滴定法)Constant current pulse time control(Coulomb electric quantity method).
Endpoint detection:Dual platinum detection electrode AC polarization method.
Terminal judgment:Drift stability determination, or setting measurement time.
Measurement of Titration by Electrolysis Method
measurement range:Reagent titration degree15(According to the prescribed injection volume).
accuracy:less than1.0%(n=3)(injection volume0.5mL, 滴定度3mg/mL).
Control method:Constant voltage control.
Endpoint detection:Dual platinum detection electrode AC polarization method.
Terminal judgment:Drift stability determination.
Main Control Unit
number of methods: 120Group measurement method.
data storage:500group data.
display function:8.4Inch color LCDtouch controlscreenCan simultaneously control four measuring unitsChinese/English/day/KoreanOperating interface.
GLPcertification:Operator NameandUse group managementdropfixedDegree measurement notificationandRemaining Quantity NotificationNotice of replacement of burette pistonandNotice of titrant replacementTitration measurement recordandInspection date notificationInspection date notificationandRecord of inspection resultsDisplay operation time.
External output:RS-232C×4USB×1SS-USB×1LAN×1.
power supply:AC100240V±10%50Hz/60Hz。
power consumption:about50 watts.