The experiment should be conducted at (20 ℃ 5) ℃ and (55 ℃ 5) Conducted at an ambient temperature of ℃. Place the battery with a thermocouple connected (with the thermocouple contact fixed at the center of the battery surface) in a fume hood and a high-temperature box respectively (for 55 ℃ The battery for the 5 ℃ short-circuit test should first be placed in a high-temperature chamber at 55 ℃; Maintain at 5 ℃ for 1.5 hours to 2 hours, short-circuit the positive and negative poles, and ensure that the total resistance of the circuit is not greater than 100m Ω. The test is terminated when the battery load voltage is less than 0.1V and the surface temperature of the battery returns to no more than 10 ℃ above the ambient temperature.
short circuit-Temperature controlled short-circuit testing machine
1requirements
The experiment should be conducted separately(20℃±5)℃ and(55℃±5)Conducted at an ambient temperature of ℃. Will be connectedthermocoupleThe battery(hot
The contact of the thermocouple is fixed at the center of the large surface of the battery)Place them separately in the fume hood and high-temperature box(conduct55℃±5℃of
The battery for short-circuit test should be first placed in a high-temperature chamber55℃±5℃Keep it down1.5h~2h)Short circuit its positive and negative poles, total power of the circuit
Resistance not greater than100mΩ. Until the battery load voltage is less than0.1VAnd the surface temperature of the battery is restored to no higher than the ambient temperature
10℃At that time, end the experiment. Each temperature test3Only battery. Internally installed recoverable temperature or overcurrent protection devices
The pool can be equipped with a resistance value that will not cause the device to short circuit the positive and negative terminals of the battery under the largest load.
2、 Purpose of the experiment:The remote control high current short-circuit device is designed based on various battery short-circuit test standards. According to the standard requirements, the short-circuit device must meet an internal resistance range far less than 100m Ω in order to obtain the maximum short-circuit current required for the test.
3、 Compliant with standards:GB/T 2900.11-1988TheMT/T 1051-2007TheSJ/T 11169-1998TheYD 1268-2003
GB/T 19521.11-2005TheGB/T 18287-2000TheSJ/T 11170-1998TheUN38.3TheUL 1642TheUL2054
QC/T 743-2006TheQC/T 744-2006TheQB/T 2502-2000TheGB/T 8897.4-2002TheYD/B 032—2009The
QB/T 2947.3-2008
four、Technical Specifications:
| model | BE-1000W |
| short-circuit current |
500AThe1000Aoptional
|
| DC response time | ≤5μs |
| Device internal resistance |
≤100mΩ, can customize resistors according to requirements
|
| Remote control distance | No solid barrier 7m |
| temperature range |
0~80℃(Adjustable)
|
| temperature deviation |
±2℃
|
| Short circuit time |
1~9999SOr 1M~99H59M can be set
|
|
Short circuit machinesize
|
W600 X D800 X H970mm
|
| Dimensions inside explosion-proof box |
W500 X D550 X H500 mm(You can choose a larger inner box)
|
| compressor | Taikang, France |
| power supply | AC 220V 50HZ communication |
| power | 4.0KW |
| weight | About 220Kg |