Just like the filament in an incandescent lamp, the filament in an ion source will eventually burn out. The following operation can avoid premature filament
Damage:
When setting the acquisition parameters, set the solvent delay so that the analyzer is turned off when the solvent peak flows out
When the software prompts "Override solution delay at the beginning of a run" at the beginning of the run
Cancel solvent delay), be sure to choose "No"
Higher emission current will reduce the lifespan of the filament
If you are controlling MSD on the parameter editing screen, be sure to select "MS Off" before changing any filament parameters
| MSD filament ordering information: | |||||
| Instructions | 5975 series | 5975T series | 5973 series | 5972 series | 5971 series |
| Filament assembly (EI) | G3170-60050 | G3170-60050 | G3170-60050 | 05971-60140 | |
| Filament Component (CI) | G1099-80053 | G1099-80053 | |||
| Micro ion vacuum gauge | G3170-80001 | ||||
| A triode gauge for measuring vacuum | 0960-0897 | ||||
| Ionic gauge controller | G3880-80010 | ||||
| Ionic regulation | G3880-80011 |
Swapping between two filaments every three months is very useful because when one filament fails, you will know that the other filament is also about to fail. So you can replace both filaments at the same time. Since the GC/MS system has already been emptied, it is a good idea to replace other spare parts while replacing the filament.