Product Description: The laser gas analysis analyzer is a high-precision, low-temperature drift trace water analysis instrument based on tunable laser absorption spectroscopy technology. Through a gas absorption cell with multiple reflections, it can provide an effective optical path of nearly 3 meters and achieve a detection limit of nearly 0.1 ppm. The integrated design of laser, detector and air chamber greatly reduces the interference of air background on measurement. Meanwhile, the compact absorption tank structure and anti adsorption treatment on the inner wall of the gas chamber effectively improve the response speed of trace water measurement.

Product Details

Basic Introduction

The laser gas analysis analyzer is a high-precision, low-temperature drift trace water analysis instrument based on tunable laser absorption spectroscopy technology. Through a gas absorption cell with multiple reflections, it can provide an effective optical path of nearly 3 meters and achieve a detection limit of nearly 0.1 ppm. The integrated design of laser, detector and air chamber greatly reduces the interference of air background on measurement. Meanwhile, the compact absorption tank structure and anti adsorption treatment on the inner wall of the gas chamber effectively improve the response speed of trace water measurement.

Application field

Natural gas, petrochemicals, steel smelting SF6、 Analysis and control of water dew point in fields such as lithium batteries.

Product Parameters


Measurement objectH2O (dew point)

Range0-10ppm to 0-2000ppm optional, expandable to 20000ppm

Measurement principleTunable Laser Absorption Spectroscopy (TDLAS)

Preheating time≤2min

Gas supply methodMobile or plug-in

ResolutionDisplay 0.1ppm (actually better than 0.1ppm)

Detection limit0.1ppm

Traffic volume0.5~2L/min

Response timeT90 ≤ 3min (from air to 10ppm)

Repeatability±1%FS

Linear error±1%FS

Zero drift± 2% FS/12 months

Range drift± 2% FS/12 months

Operating temperature-10℃~50℃

Storage temperature-20℃~60℃

Digital outputRS485 protocol

Working power supply24V DC

Analog output4~20mA


Technical Features

Integrated design of laser, detector and multiple reflection chamber, without air background interference.

Multiple reflection pools increase the detection limit of trace water, reaching up to 0.1ppm.

Compact chamber structure and anti adsorption treatment on the inner wall improve response speed.